Why the study?
Does deformation-tracking using CMR cine imaging provide comparable diagnostic performance for circumferential strain assessment compared to DENSE and feature-tracking in healthy adults?
Does deformation-tracking using CMR cine imaging provide comparable diagnostic performance for circumferential strain assessment compared to DENSE and feature-tracking in healthy adults?
Deformation-tracking using standard CMR cine imaging provides reproducible circumferential strain assessment comparable to DENSE, potentially eliminating the need for additional bespoke strain sequences and reducing scan time.
Answers are grounded in this paper and cite their sources.
Deformation-tracking may enable reproducible strain assessment on standard cine CMR; leaves open prospective validation before replacing dedicated sequences.
We developed a novel method for tracking myocardial deformation using cardiac magnetic resonance (CMR) cine imaging. We hypothesised that circumferential strain using deformation-tracking has comparable diagnostic performance to a validated method (Displacement Encoding with Stimulated Echoes- DENSE) and potentially diagnostically superior to an established cine-strain method (feature-tracking). 81 healthy adults (44.6 ± 17.7 years old, 47% male), without any history of cardiovascular disease, underwent CMR at 1.5 T including cine, DENSE, and late gadolinium enhancement in subjects >45 years. Acquisitions were divided into 6 segments, and global and segmental peak circumferential strain were derived and analysed by age and sex. Peak circumferential strain differed between the 3 groups (DENSE: -19.4 ± 4.8%; deformation-tracking: -16.8 ± 2.4%; feature-tracking: -28.7 ± 4.8%) (ANOVA with Tukey post-hoc, F-value 279.93, p < 0.01). DENSE and deformation-tracking had better reproducibility than feature-tracking. Intra-class correlation co-efficient was >0.90. Larger magnitudes of strain were detected in women using deformation-tracking and DENSE, but not feature-tracking. Compared with a reference method (DENSE), deformation-tracking using cine imaging has similar diagnostic performance for circumferential strain assessment in healthy individuals. Deformation-tracking could potentially obviate the need for bespoke strain sequences, reducing scanning time and is more reproducible than feature-tracking.
No takes yet. Share an insight, caveat, or question.
Mangion et al. (2016) studied this question.