Key result
Prolonged vasovagal asystole disables S-ICD SMART Pass, triggering T-wave oversensing and inappropriate shocks.
Why the study?
Given the advantages of S-ICD in younger populations, understanding clinical factors that may cause inappropriate device function is critical.
Case Report (n=1)
No
Clinicians managing patients with S-ICDs must recognize that prolonged bradycardia or asystole (>10 seconds), such as from vasovagal syncope, automatically disables the SMART Pass filter and predisposes patients to T-wave oversensing and inappropriate shocks.
Answers are grounded in this paper and cite their sources.
Prolonged asystole may disable S-ICD SMART Pass, risking shocks in syncope; case extends device interaction data but leaves confirmation open.
Subcutaneous ICD (S-ICD) has several advantages in younger patient populations due to the absence of intracardiac and transvenous hardware. It is therefore critical for the modern-day physician to understand clinical factors which may lead to inappropriate device function in these ICD technologies.
No takes yet. Share an insight, caveat, or question.
Miranda et al. (2026) conducted a case report in Surgically corrected dextro-transposition of the great arteries (d-TGA) and ventriculoseptal defect (VSD) with inducible ventricular tachycardia and vagally mediated syncope (n=1). Subcutaneous implantable cardioverter-defibrillator (S-ICD) with SMART Pass filter and vagal trigger avoidance counseling was evaluated on Inappropriate S-ICD shocks due to T-wave oversensing following SMART Pass filter autodisabling from prolonged asystole (>10 seconds). Vagally mediated syncope with prolonged asystole (>10 seconds) caused automatic disabling of the S-ICD SMART Pass filter, leading to T-wave oversensing and inappropriate shocks.
Synapse has enriched 10 related papers. Consider them for comparative context: