Key result
Patient-specific 3D-printed bench testing enables successful guiding catheter selection and PCI in anomalous RCA.
Why the study?
Percutaneous coronary intervention of anomalous aortic origin of a coronary artery is technically challenging due to difficult engagement and poor support, and conventional catheter selection often fails.
Does preprocedural bench testing with a patient-specific 1:1 scale 3D-printed coronary model facilitate successful guiding catheter selection and PCI in a patient with an anomalous right coronary artery originating from the left sinus after failed conventional engagement attempts?
Case Report (n=1)
No
Does preprocedural bench testing with a patient-specific 1:1 scale 3D-printed coronary model facilitate successful guiding catheter selection and PCI in a patient with an anomalous right coronary artery originating from the left sinus after failed conventional engagement attempts?
Patient-specific 3D-printed coronary modeling enables physical bench testing and precise guiding catheter selection to facilitate successful PCI in complex anomalous aortic origin of the right coronary artery after failed conventional engagement.
Answers are grounded in this paper and cite their sources.
May aid catheter selection in anomalous RCA after failed engagement; hypothesis-generating for broader 3D-model adoption.
BACKGROUND: Percutaneous coronary intervention (PCI) of an anomalous aortic origin of a coronary artery (AAOCA) is technically challenging given difficult engagement and poor support. While coronary computed tomography angiography aids planning, it lacks the tactile simulation provided by patient-specific 3-dimensional (3D) modeling. CASE SUMMARY: A 58-year-old woman presented with recurrent angina and an anomalous right coronary artery originating from the left sinus. After repeated failures using conventional catheter selection, a 1:1 scale 3D-printed model was used for bench testing. This enabled the selection of an EBU 3.0 guiding catheter, leading to a successful PCI with 2 drug-eluting stents. DISCUSSION: This case demonstrates that 3D bench testing allows operators to anticipate support limitations and select optimal catheter geometry, potentially reducing procedure time and radiation. It highlights a structured alternative to empirical "trial-and-error" strategies in complex congenital anatomy. TAKE-HOME MESSAGE: Patient-specific 3D modeling facilitates successful PCI in AAOCA by enabling precise preprocedural simulation and optimal guiding catheter selection.
No takes yet. Share an insight, caveat, or question.
Jossart et al. (2026) conducted a case report in Anomalous aortic origin of the right coronary artery (AAORCA) with recurrent angina and severe mid-to-distal stenoses (n=1). Patient-specific 1:1 scale 3D-printed coronary model bench testing to guide PCI vs. Conventional empirical guiding catheter selection (prior failed attempts) was evaluated on Procedural success (selective engagement of anomalous RCA, stent implantation, and final TIMI flow grade 3) and freedom from angina at 3 months. Patient-specific 1:1 scale 3D-printed bench testing enabled selection of an EBU 3.0 guiding catheter, achieving successful PCI with 2 drug-eluting stents and TIMI 3 flow in an anomalous RCA.