Why the study?
Competitive atrial pacing during sensor-driven pacing competing with intrinsic atrial events falling in PVARP can cause proarrhythmia, requiring potential programming solutions or corrective algorithms.
Does inactivation of the activity sensor or change to atrial-based pacing modes prevent device-induced atrial proarrhythmia in patients with competitive atrial pacing?
Does inactivation of the activity sensor or change to atrial-based pacing modes prevent device-induced atrial proarrhythmia in patients with competitive atrial pacing?
Inactivating the activity sensor or switching to AAI/R modes can prevent competitive atrial pacing-induced proarrhythmia, highlighting the need for corrective algorithms.
Answers are grounded in this paper and cite their sources.
May support sensor inactivation or AAI/R reprogramming to avoid device-induced atrial proarrhythmia; leaves open prospective validation.
CAP predominantly occurs during sensor-driven atrial pacing that competes with intrinsic atrial events falling in PVARP. Inactivation of the activity sensor or change to atrial-based pacing modes (AAI/R) appears to effectively prevent induction of device-induced atrial proarrhythmia. Ultimately, a corrective algorithm is needed to avoid CAP-induced proarrhythmia.
No takes yet. Share an insight, caveat, or question.
Wass et al. (2020) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: