Why the study?
Precise positioning of transcatheter heart valves inside surgical heart valves is crucial for optimal functional results, but the impact of implantation depth required systematic investigation.
Does the implantation depth of a transcatheter heart valve inside a surgical heart valve affect hydrodynamic performance in an in vitro model?
Does the implantation depth of a transcatheter heart valve inside a surgical heart valve affect hydrodynamic performance in an in vitro model?
The optimal implantation depth for hydrodynamic performance of the Evolut PRO transcatheter heart valve in a valve-in-valve procedure differs depending on the specific surgical heart valve model.
Answers are grounded in this paper and cite their sources.
Optimal depths vary by surgical valve in vitro; leaves open translation to clinical ViV outcomes.
OBJECTIVES: Transcatheter aortic valve-in-valve (ViV) procedures are increasingly performed for the treatment of degenerated surgical aortic valves with a high risk for a redo operation. For an optimal functional result, precise positioning of the transcatheter heart valve (THV) inside the SHV is crucial. The aim of this study was to systematically investigate the impact of implantation depth on the functional result after a ViV procedure in a standardized in vitro setting. METHODS: A THV 23 mm (Evolut PRO) and 3 SHV 21 mm (Perimount Magna Ease, Trifecta and Hancock II) were used for hydrodynamic testing with a constant heartbeat 64/min and a range of 55-105 ml of stroke volume in 5 different positions of the THV. The following parameters were analysed: mean pressure gradient (MPG), effective orifice area (EOA), geometric orifice area, minimal internal diameter and pin-wheeling index. RESULTS: MPG and EOA differed significantly regarding the position of the THV in the same SHV. The highest EOA and the lowest MPG were recorded for Evolut PRO with significance for both parameters in Hancock II at 4 vs 5 mm (P < 0.001), in Magna Ease at 2 mm (vs 3 mm and vs 6 mm, P < 0.001) and in Trifecta at 4 mm (vs 5 and 6 mm, P < 0.001). Leaflet coadaptation, minimal internal diameter and maximal geometric orifice area of the same TAV differ regarding the position of the TAV. CONCLUSIONS: The optimal position for hydrodynamic performance of the THV as ViV differs among specific SHV models. The findings may be useful for planning a ViV procedure using the Evolut PRO THV.
No takes yet. Share an insight, caveat, or question.
Sadat et al. (2022) studied this question.