Key result
In-vitro measurements of 12 bileaflet valve samples accurately determined the dependence of the discharge coefficient on internal diameter and flow rate for use in the Gorlin formula.
This in-vitro study accurately determined the discharge coefficient for bileaflet valves, which can improve the accuracy of Doppler-based valvular orifice measurements using the Gorlin formula.
Answers are grounded in this paper and cite their sources.
May refine Gorlin constants for bileaflet valves; leaves open in-vivo validation before clinical use.
The measurement of a cardiac valve's area is a common procedure, usually performed with noninvasive, Doppler-based techniques. Such measurements are not, however, without problems: a potential source of errors is the value of a valve's discharge coefficient. In-vitro pressure and flow measurements relative to the bileaflet valve of four brands were performed. A total of 12 valve samples was studied to cover the entire range of valve sizing. The data were used in the Gorlin formula for valve area measurements, and the dependence of the discharge coefficient on the internal diameter of the valve and the flow rate was accurately determined. The reported results can be used in a great number of follow-up clinical assessments to improve the accuracy of valvular orifice measurements.
No takes yet. Share an insight, caveat, or question.
Grigioni et al. (2005) studied Cardiac valve area measurement (n=12). In-vitro pressure and flow measurements was evaluated on Dependence of the discharge coefficient on the internal diameter of the valve and the flow rate. In-vitro measurements of 12 bileaflet valve samples accurately determined the dependence of the discharge coefficient on internal diameter and flow rate for use in the Gorlin formula.
Synapse has enriched 10 related papers. Consider them for comparative context: