Key result
EOA by continuity equation uniquely predicts postoperative sPAP among prosthesis-patient mismatch definitions after MVR.
Why the study?
Does the method of calculating effective orifice area (CE vs PHT vs reference) affect the diagnosed incidence of prosthesis-patient mismatch and its correlation with hemodynamics in patients after mitral valve replacement?
Observational (n=166)
No
Does the method of calculating effective orifice area (CE vs PHT vs reference) affect the diagnosed incidence of prosthesis-patient mismatch and its correlation with hemodynamics in patients after mitral valve replacement?
Answers are grounded in this paper and cite their sources.
The continuity equation is the most reliable method for calculating effective orifice area to define prosthesis-patient mismatch after mitral valve replacement, as it uniquely predicts postoperative hemodynamic parameters.
PURPOSE: The incidence of prosthesis-patient mismatch (PPM) after mitral valve replacement (MVR) has been reported to vary. The purpose of the current study was to investigate incidence of PPM according to the different methods of calculating effective orifice area (EOA), including the continuity equation (CE), pressure half time (PHT) method and use of reference EOA, and to compare these with various echocardiographic variables. MATERIALS AND METHODS: We retrospectively reviewed 166 individuals who received isolated MVR due to rheumatic mitral stenosis and had postoperative echocardiography performed between 12 and 60 months after MVR. EOA was determined by CE (EOA(CE)) and PHT using Doppler echocardiography. Reference EOA was determined from the literature or values offered by the manufacturer. Indexed EOA was used to define PPM as present if ≤1.2 cm²/m². RESULTS: Prevalence of PPM was different depending on the methods used to calculate EOA, ranging from 7% in PHT method to 49% in referred EOA method to 62% in CE methods. The intraclass correlation coefficient was low between the methods. PPM was associated with raised trans-prosthetic pressure, only when calculated by CE (p=0.021). Indexed EOA(CE) was the only predictor of postoperative systolic pulmonary artery (PA) pressure, even after adjusting for age, preoperative systolic PA pressure and postoperative left atrial volume index (p<0.001). CONCLUSION: Prevalence of mitral PPM varied according to the methods used to calculate EOA in patients with mitral stenosis after MVR. Among the various methods used to define PPM, EOA(CE) was the only predictor of postoperative hemodynamic parameters.
No takes yet. Share an insight, caveat, or question.
Cho et al. (2016) conducted an observational in Rheumatic mitral stenosis after mitral valve replacement (n=166). Effective orifice area calculation by continuity equation (EOACE) vs. Pressure half time (PHT) method and reference EOA was evaluated on Postoperative systolic pulmonary artery pressure (β -0.37, 95% CI -12.2 to -5.8, p=<0.001). Among various methods used to define prosthesis-patient mismatch after mitral valve replacement, the effective orifice area calculated by the continuity equation was the only independent predictor of postoperative systolic pulmonary artery pressure.
Synapse has enriched 10 related papers. Consider them for comparative context: