Key result
MOLLI sampling achieves high T1 mapping accuracy with an error rate under ~6%.
Why the study?
The study aimed to clarify the accuracy and variability of heart rate variability effects on T1 relaxation time values acquired using different MOLLI sampling schemes.
Does the 5(3)3 MOLLI sampling scheme improve accuracy and reduce heart rate variability effects in myocardial T1 mapping compared to other schemes in a phantom model?
Does the 5(3)3 MOLLI sampling scheme improve accuracy and reduce heart rate variability effects in myocardial T1 mapping compared to other schemes in a phantom model?
The 5(3)3 MOLLI sampling scheme provides highly accurate T1 measurements and is robust against heart rate variability in phantom models.
Answers are grounded in this paper and cite their sources.
May support 5(3)3 MOLLI for accurate T1 mapping in phantoms; leaves open clinical translation amid heart rate variability.
This study aimed to clarify the accuracy and variability of heart rate variability effects on T1 relaxation time values acquired using different sampling schemes of the modified Look-Locker inversion recovery (MOLLI) method. Five MOLLI sampling schemes-5(3)3, 5s(3s)3s, 4(1)3(1)2, 4s(1s)3s(1s)2s, and 3(3)3(3)5-were compared using a simulated electrocardiogram signal in a phantom experiment. The 5(3)3 and 5s(3s)3s schemes demonstrated high accuracy across the range of native and contrast-enhanced T1 values, with correlation coefficients exceeding 0.95 compared with the reference T1. Regarding heart rate variability, the 5(3)3 scheme showed minimal variation in the measured value across different heart rates, regardless of the T1 value, with an error rate below 6% relative to the reference T1. Therefore, the 5(3)3 scheme can achieve highly accurate measurements over a wide T1 range and is robust against heart rate variability.
No takes yet. Share an insight, caveat, or question.
Endo et al. (2025) studied Myocardial T1 mapping (Phantom study). MOLLI sampling schemes (5(3)3, 5s(3s)3s, 4(1)3(1)2, 4s(1s)3s(1s)2s, and 3(3)3(3)5) vs. Reference T1 values was evaluated on Accuracy and variability of T1 relaxation time values across different heart rates. The 5(3)3 MOLLI sampling scheme demonstrated high accuracy across a wide range of T1 values with correlation coefficients exceeding 0.95 and an error rate below 6% relative to reference T1.