Key result
A new automated model for estimating mitral regurgitant orifice diameter and volume showed close agreement with reference values in an in vitro flow model.
A novel automated echocardiographic method using normalized centerline velocity distribution accurately estimates mitral regurgitant orifice diameter and volume in an in vitro model, overcoming limitations of the PISA technique.
Answers are grounded in this paper and cite their sources.
Hypothesis-generating for automated mitral regurgitation quantification; clinical validation required before any use.
Previous echocardiographic techniques for quantifying valvular regurgitation (PISA) are limited by factors including uncertainties in orifice location and hemispheric convergence assumption. Using computational fluid dynamics simulations, we developed a new model for the estimation of orifice diameter and regurgitant volume without the aforementioned assumptions of the PISA technique. Using experimental data obtained from the in vitro flow model we successfully validated our new model. The model output (y) and reference (x) values were in close agreement (y = 0.95x + 0.38, r = 0.96, error = 1.68 +/- 7.54% for the orifice diameter and y = 1.18x - 4.72, r = 0.93, error = 6.48 +/- 16.81% for the regurgitant volume).
No takes yet. Share an insight, caveat, or question.
Deserranno et al. (2003) studied Mitral Regurgitation. New automated method using Color M-mode and Continuous Wave Doppler vs. Reference values was evaluated on Orifice diameter and regurgitant volume (r = 0.96 (diameter), r = 0.93 (volume)). A new automated model for estimating mitral regurgitant orifice diameter and volume showed close agreement with reference values in an in vitro flow model.
Synapse has enriched 10 related papers. Consider them for comparative context: