Key result
An automated algorithm using digital Doppler velocity maps to calculate flow rate showed good agreement with actual flow rates during in vitro testing.
An automated algorithm using digital Doppler velocity maps accurately calculates flow rate in an in vitro model, potentially improving the quantification of valvular regurgitation.
Answers are grounded in this paper and cite their sources.
May support automated flow quantification research; leaves open in vivo and clinical validation.
To automate quantification of valvular regurgitation, the authors developed an algorithm which uses the full proximal convergence zone (output as a digital map) to automatically define the orifice center and calculate the flow rate. For a user defined region of interest, the actual velocity field is compared with the expected velocity map generated by the function nu =(Qcos psi /2 pi r/sup 2/), where r is the distance from the test center, psi the angle between the converging velocity and the ultrasound beam, and Q the flow rate which is being solved for. The optimal orifice center is that point which maximizes the correlation between observed and predicted velocity. In vitro testing revealed good agreement with flow rate: r=0.98, Delta Q=-3.7+or-15.8 cm/sup 3//sec.< <ETX xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">></ETX>
No takes yet. Share an insight, caveat, or question.
Thomas et al. (2002) studied valvular regurgitation. Automated algorithm using digital Doppler velocity maps was evaluated on Flow rate agreement (r=0.98). An automated algorithm using digital Doppler velocity maps to calculate flow rate showed good agreement with actual flow rates during in vitro testing.
Synapse has enriched 10 related papers. Consider them for comparative context: