Key result
Epicardial LV electrodes match endocardial stimulation but are linked to ~15% higher absolute mortality.
Why the study?
Does epicardial left ventricular lead implantation provide similar electrical performance, clinical response, and safety compared to transvenous implantation in patients receiving CRT?
Case-Control (n=184)
Yes
Does epicardial left ventricular lead implantation provide similar electrical performance, clinical response, and safety compared to transvenous implantation in patients receiving CRT?
Absolute Event Rate: 98.1% vs 96.6%
p-value: p=0.62
While epicardial LV leads provide effective resynchronization comparable to transvenous leads, they require higher pacing energy and are associated with higher mortality, likely reflecting a sicker patient population requiring surgical intervention.
Supports caution with epicardial LV leads in CRT; leaves open whether higher mortality reflects selection bias or true risk.
No abstract yet. The journal hasn't published one we can show.
No takes yet. Share an insight, caveat, or question.
Fedorco et al. (2017) conducted a case-control in Heart failure indicated for cardiac resynchronization therapy (n=184). Epicardial left ventricular electrode implantation vs. Endovascular (transvenous) left ventricular electrode implantation was evaluated on Effective left ventricular stimulation at 3-year follow-up (p=0.62). Epicardial left ventricular electrodes provided effective stimulation comparable to endocardial electrodes at 3 years (98.1% vs 96.6%), but were associated with significantly higher 3-year mortality (21.7% vs 6.5%).