Interaction between two endocardial defibrillation leads was identified as a rare cause of electrical noise and oversensing during an ICD reimplantation procedure in a 68-year-old man.
Case Report (n=1)
During ICD reimplantation, if a nonfunctional lead cannot be extracted, the new lead should be implanted away from and not parallel to the old one to prevent electrical noise and oversensing.
Answers are grounded in this paper and cite their sources.
Alerts clinicians to rare lead interaction in unexplained ICD noise with retained leads; case report leaves open need for validation before practice change.
A variety of etiologies can cause erroneous detection in patients with implantable cardioverter defibrillator (ICDs). Interaction between two endocardial leads is rare and uncommon in causing electrical noise. During a reimplantation procedure of an ICD system in a 68-year-old man, additional electrical signals could be detected. The interaction between two endocardial defibrillation leads was identified as the cause of sensing problems. When it is not possible to extract the nonfunctional endocardial lead during implantation of the new electrode, it should be implanted away and not in parallel from the old one to avoid interaction between them. A variety of etiologies can cause erroneous detection in patiens with ICD. Interaction between two endocardial leads is rare and uncommon to cause electrical noise.
No takes yet. Share an insight, caveat, or question.
Gardas et al. (2006) conducted a case report in Implantable cardioverter-defibrillator (ICD) oversensing (n=1). ICD reimplantation was evaluated. Interaction between two endocardial defibrillation leads was identified as a rare cause of electrical noise and oversensing during an ICD reimplantation procedure in a 68-year-old man.