Key result
High frequency (40 MHz) intravascular ultrasound identified lipid/necrotic pools in atherosclerotic plaques with 65% sensitivity and 95% specificity compared to histomorphology.
Why the study?
Does high frequency (40 MHz) intravascular ultrasound accurately identify lipid/necrotic pools in human atherosclerotic plaques compared to histomorphology?
Does high frequency (40 MHz) intravascular ultrasound accurately identify lipid/necrotic pools in human atherosclerotic plaques compared to histomorphology?
High frequency (40 MHz) IVUS demonstrates good specificity (95%) but moderate sensitivity (65%) for identifying lipid/necrotic pools in atherosclerotic plaques in vitro.
Caution advised against clinical adoption; leaves open human in vivo validation of 40 MHz IVUS.
OBJECTIVE To test the efficacy of high frequency intravascular ultrasound (IVUS) transducers in identifying lipid/necrotic pools in atherosclerotic plaques. METHODS 40 MHz transducers were used for in vitro IVUS assessment of 12 arterial segments (10 coronary and two carotid arteries, dissected from five different necropsy cases). IVUS acquisition was performed at 0.5 mm/s after ligature of the branching points to generate a closed system. Lipid/necrotic areas were defined by IVUS as large echolucent intraplaque areas surrounded by tissue with higher echodensity. To obtain histopathological sections corresponding to IVUS cross sections, vessels were divided into consecutive 3 mm long segments using the most distal recorded IVUS image as the starting reference. Samples were then fixed with 10% buffered formalin, processed for histopathological study, serially cut, and stained using the Movat pentacrome method. RESULTS 122 sections were analysed. Lipid pools were observed by histology in 30 sections (25%). IVUS revealed the presence of lipid pools in 19 of these sections (16%; sensitivity 65%, specificity 95%). CONCLUSIONS In vitro assessment of lipid/necrotic pools with high frequency transducers was achieved with good accuracy. This opens new perspectives for future IVUS characterisation of atherosclerotic plaques.
No takes yet. Share an insight, caveat, or question.
Prati et al. (2001) studied Atherosclerotic plaques (n=5). High frequency (40 MHz) intravascular ultrasound (IVUS) vs. Histomorphology was evaluated on Identification of lipid/necrotic pools (Sensitivity 65%, Specificity 95%). High frequency (40 MHz) intravascular ultrasound identified lipid/necrotic pools in atherosclerotic plaques with 65% sensitivity and 95% specificity compared to histomorphology.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: