Key result
Pulse contour equations using aortic pressure variables showed a good correlation with measured percent changes in stroke volume during VDD pacing in dogs.
Why the study?
Do pulse contour methods accurately estimate stroke volume changes from aortic pressure during VDD pacing in a canine model?
Do pulse contour methods accurately estimate stroke volume changes from aortic pressure during VDD pacing in a canine model?
Pulse contour methods using aortic pressure parameters correlate well with stroke volume changes at the onset of VDD pacing in a canine model.
Answers are grounded in this paper and cite their sources.
Pulse contour equations correlate with stroke volume changes in dogs; leaves open human translation during VDD pacing.
Stroke volume (SV) might be an important variable in assessing cardiac function of patients under pacing therapy. To assess the accuracy of estimating SV changes from aortic pressure during VDD pacing, we implanted electromagnetic flowmeters in the aortic root of five mongrel dogs. Three pacing sites, including right ventricle, left ventricle and both ventricles were selected and paced at 5 atrioventricular delays. Each pacing site/AV delay combination was repeated 5 times in a randomized order. The protocol alternated 5 paced beats and 15 intrinsic beats. Flow and aortic pressure data were recorded and analyzed. We evaluated five pulse contour equations relating pulse pressure (PP), ejection time (ET), diastolic time (DT) and aortic pressure area during systole over the end-diastolic pressure (Psa). A good correlation (r=0.84-0.89) was shown between percent change in SV and percent change in PP, PP*ET, Psa, Psa*ET and Psa*(1+ET/DT) over the preceding baseline. The average estimated percent change in SV also correlated with the average measured percent change in SV for each pacing site/AV delay combination. In conclusion, PP, PP*ET, Psa, Psa*ET and Psa*(1+ET/DT) may be clinically applicable for assessing SV at the start of VDD pacing and may effectively predict the effects of pacing sites and AV delays on SV.
No takes yet. Share an insight, caveat, or question.
Yu et al. (2002) studied VDD pacing (n=5). Pulse contour equations vs. Electromagnetic flowmeter measurements was evaluated on Correlation between percent change in estimated stroke volume and measured stroke volume (r=0.84-0.89). Pulse contour equations using aortic pressure variables showed a good correlation with measured percent changes in stroke volume during VDD pacing in dogs.
Synapse has enriched 8 closely related papers on similar clinical questions. Consider them for comparative context: