Does computer estimation of stroke volume from aortic pulse contour accurately measure actual stroke volume in dogs and humans?
Does computer estimation of stroke volume from aortic pulse contour accurately measure actual stroke volume in dogs and humans?
Computer estimation of stroke volume from aortic pulse contour using standard formulas is consistent but inaccurate compared to reference methods in dogs and humans.
Answers are grounded in this paper and cite their sources.
Warner's formula warrants caution for stroke volume estimation; leaves open validation of refined algorithms in larger human cohorts.
Stroke volume (SV) estimation from computer analysis of aortic pressure contour was tested in 12 anesthetized dogs and 2 humans with A-V block. The ‘actual’ SV was calculated from right-heart bypass pump flow values in 6 open-chest dogs, from heart rate (HR) and cardiac output (CO) using dye dilution in 6 closed-chest dogs and by the Fick method in the human subjects. SV changes were induced by changes in pacing rate and pump flow, infusion and withdrawal of fluids and exercise. Mean arterial pressure was altered in dogs by changes in CO and infusions of drugs.In the dogs, the method (using Warner’s formula) proved consistent (r = 0.774, p < 0.001, Sy. x = 3.82 ml) but inaccurate (slope = 0.386) in 627 comparisons. The same conclusions were reached in 11 human comparisons. The Kouchoukos formula showed similar problems. Corrections are suggested based on these results.