Key result
Simple photoplethysmogram-based AV delay optimization using signal slope was as reproducible as Finometer-based optimization.
Why the study?
Does a simple photoplethysmogram signal provide reproducible automatic AV delay optimization compared to Finometer in patients with CRT?
Does a simple photoplethysmogram signal provide reproducible automatic AV delay optimization compared to Finometer in patients with CRT?
A simple photoplethysmogram signal can provide fully automatic and reproducible AV delay optimization in CRT patients, comparable to bulkier equipment when using signal slope.
Answers are grounded in this paper and cite their sources.
May simplify AV optimization in CRT; leaves open whether it improves outcomes versus standard methods.
BACKGROUND: Hemodynamic optimization of cardiac resynchronization therapy (CRT) can be achieved reproducibly and--with bulky, nonimplantable equipment--noninvasively. We explored whether a simple photoplethysmogram signal might be used instead. METHOD: Twenty patients (age 65 ± 12) with CRT underwent automatic atrioventricular (AV) delay optimization, using a multiple-transitions protocol, at two atrially paced heart rates: just above sinus rate ("slow ApVp," 77 ± 11 beats per minute [bpm]) and 100 bpm ("fast ApVp"). We then retested to assess short-term reproducibility. RESULTS: All 80 optimizations identified an optimum (correctly oriented parabola). At 100 bpm, the simple photoplethysmogram had wider scatter between repeat optimizations than did Finometer: standard deviation of difference (SDD) 22 ms versus 14 ms, respectively, P = 0.028. The simple photoplethysmogram improved in reproducibility when slope (instead of peak) of its signal was used for optimization, becoming as reproducible as Finometer (SDD 14 ms vs 14 ms, P = 0.50). At slow heart rate, reproducibility of simple photoplethysmogram-based optimization worsened from 14 to 22 ms (P = 0.028), and Finometer-based optimization from 14 to 26 ms (P = 0.005). Increasing the number of replicates averaged improved reproducibility. For example, SDD of simple photoplethysmogram optimization (using peak) fell from 62 ms with two replicates to 22 ms with eight replicates (P < 0.0001). At 100 bpm, the eight-replicate protocol takes ∼12 minutes. CONCLUSIONS: A 12-minute protocol of simple photoplethysmographic AV optimization can be processed fully automatically. Blinded test-retest reproducibility of the optimum AV is good and improves with more replicates. If benefits to some patients are not to be neutralized by harm to others, endpoint studies should first test check narrowness of "within-patient error bars."
No takes yet. Share an insight, caveat, or question.
Kyriacou et al. (2012) studied Cardiac resynchronization therapy (CRT) (n=20). Simple photoplethysmogram-based AV delay optimization vs. Finometer-based optimization was evaluated on Standard deviation of difference (SDD) between repeat optimizations at 100 bpm (using peak signal) (p=0.028). Simple photoplethysmogram-based AV delay optimization using signal slope was as reproducible as Finometer-based optimization.