Key result
CONTEC08C automated upper-arm sphygmomanometer meets ISO validation criteria for SBP and DBP accuracy.
Why the study?
This study was conducted to assess the accuracy of the CONTEC08C automated oscillometric upper-arm sphygmomanometer in adults for clinical use in accordance with the ISO 81060-2:2018 standard.
Does the CONTEC08C automated oscillometric upper-arm sphygmomanometer accurately measure blood pressure in adults compared to reference measurements?
Cross-Sectional (n=85)
Does the CONTEC08C automated oscillometric upper-arm sphygmomanometer accurately measure blood pressure in adults compared to reference measurements?
The CONTEC08C oscillometric upper-arm blood pressure monitor meets ISO 81060-2:2018 accuracy requirements and is recommended for clinical use in adults.
Answers are grounded in this paper and cite their sources.
Extends device validation data for CONTEC08C; leaves open generalizability across diverse adult populations and settings.
This study aimed to assess the accuracy of the CONTEC08C automated oscillometric upper-arm sphygmomanometer in adults for clinical use in accordance with the ISO 81060-2:2018 standard. A total of 90 normotensive and hypertensive participants aged 18-75 years were recruited, and 85 were analyzed to evaluate the accuracy of the CONTEC08C oscillometric upper-arm sphygmomanometer using the same arm sequential validation method and a standardized protocol. The mean ± standard deviation (SD) of the differences between the test device and reference blood pressure (BP) measurements was 0.5 ± 3 mmHg for SBP and 0.4 ± 3.5 mmHg for DBP, meeting validation criterion 1. For validation criterion 2, the SD of the averaged differences between the test device and reference BP measurements per subject was 1.81/2.41 mmHg (systolic/diastolic). The CONTEC08C oscillometric upper-arm blood pressure monitor fulfilled the requirements of the ISO 81060-2:2018 standard in adults in the clinical setting and can therefore be recommended for safe clinical use.
No takes yet. Share an insight, caveat, or question.
Chen et al. (2025) conducted a cross-sectional in Normotension and hypertension (n=85). CONTEC08C automated oscillometric upper-arm sphygmomanometer vs. Reference blood pressure measurements was evaluated on Accuracy according to ISO 81060-2:2018 standard (mean ± SD of differences between test device and reference). The CONTEC08C automated oscillometric upper-arm sphygmomanometer met ISO 81060-2:2018 validation criteria with mean differences of 0.5 ± 3 mmHg for SBP and 0.4 ± 3.5 mmHg for DBP.