Key result
Severe prosthesis oversizing was associated with contained rupture of the aortic root in patients undergoing balloon-expandable TAVI.
Why the study?
Does prosthesis oversizing increase the risk of contained rupture of the aortic root in patients undergoing balloon-expandable TAVI?
Observational (n=72)
Does prosthesis oversizing increase the risk of contained rupture of the aortic root in patients undergoing balloon-expandable TAVI?
Severe prosthesis oversizing during balloon-expandable TAVI is strongly associated with contained rupture of the aortic root, highlighting the importance of accurate CT-based sizing.
Answers are grounded in this paper and cite their sources.
Severe oversizing may heighten contained rupture risk in balloon-expandable TAVI; leaves open optimal CT sizing thresholds.
BACKGROUND: To retrospectively investigate the potential cause of contained rupture of the aortic root in balloon-expandable transcatheter aortic valve implantation (TAVI) by means of pre- and postinterventional multislice computed tomography. METHODS AND RESULTS: Seventy-two patients (mean age 82±7 years, mean aortic valve area 0.69±0.19 cm(2)) underwent balloon-expandable TAVI using the EdwardsSAPIEN Transcatheter Heart Valve (23 mm, n=19; 26 mm, n=50; 29 mm, n=3). Aortic annulus dimensions were quantified by multislice computed tomography-based cross-sectional area assessment and average diameter calculation (CAAD) before and after TAVI. Post-TAVI multislice computed tomography data sets were available in 65 patients; contained aortic root rupture was diagnosed in 3 patients. Pre-TAVI CAAD was 23.1±1.8 mm; post-TAVI CAAD was 22.9±1.3 mm. Median relative change in CAAD pre- and post-TAVI was -0.5% (interquartile range, 3.6%). Relative increase of 5% to 10% was observed in 4 patients (1 with contained rupture), relative increase >10% in 2 patients, both with contained rupture. Mean relative oversizing, calculated as the relative difference in diameter between pre-TAVI CAAD and nominal diameter of the selected prosthesis, was 9.8%±7.8%. Relative oversizing was significantly higher in patients with contained rupture compared with patients without contained rupture (24.6%±5.4% versus 9.1%±6.6%; P<0.001). Relative oversizing ≥20% occurred in 6 patients (3 with contained rupture). CONCLUSIONS: Contained rupture of the aortic root in balloon-expandable TAVI is associated with severe prosthesis oversizing. Multislice computed tomography-based assessment of aortic annulus dimension in conjunction with adapted sizing guidelines may reduce the incidence of severe oversizing.
No takes yet. Share an insight, caveat, or question.
Blanke et al. (2012) conducted an observational in Aortic stenosis (n=72). Balloon-expandable TAVI (Edwards SAPIEN) was evaluated on Relative prosthesis oversizing in patients with vs without contained aortic root rupture (p=<0.001). Severe prosthesis oversizing was associated with contained rupture of the aortic root in patients undergoing balloon-expandable TAVI.
Synapse has enriched 10 related papers. Consider them for comparative context: