Key result
Noninvasive 3D cardiac electrical imaging accurately localized the initiation site of activation with an average error of 7.0+/-1.1 mm for RV and 6.6+/-1.9 mm for LV sites in a swine model.
Why the study?
Does noninvasive 3D cardiac electrical imaging accurately estimate the initiation site and activation sequence compared to intracavitary noncontact mapping in a swine model?
Does noninvasive 3D cardiac electrical imaging accurately estimate the initiation site and activation sequence compared to intracavitary noncontact mapping in a swine model?
Noninvasive 3D cardiac electrical imaging can accurately localize the initiation site and estimate cardiac activation sequences in an in vivo swine model.
Supports preclinical validation in swine; leaves open human translation before clinical adoption.
A noninvasive three-dimensional (3D) cardiac electrical imaging (3DCEI) approach, which can estimate the location of the initiation site (IS) of activation and the resultant 3D activation sequence (AS) from body surface potential maps (BSPMs), was validated in an intact large mammalian model (swine) during acute ventricular pacing. Body surface potential mapping and intracavitary noncontact mapping (NCM) were performed simultaneously during pacing from both right ventricular (RV) sites (intramural) and left ventricular (LV) sites (endocardial). Subsequent 3DCEI analyses were performed on the measured BSPMs. In total, 5 RV and 5 LV sites from control and heart failure animals were paced. The averaged localization error of the RV and LV sites were 7.0+/-1.1 mm and 6.6+/-1.9 mm, respectively. The endocardial ASs as a subset of the estimated 3D ASs by 3DCEI were consistent with those reconstructed from the NCM system. The present experimental results demonstrate that the noninvasive 3DCEI approach can localize the initiation site and estimate cardiac activation sequence with good accuracy in an in vivo setting, under control, paced and/or diseased conditions.
No takes yet. Share an insight, caveat, or question.
Liu et al. (2008) studied Ventricular pacing in control and heart failure models. Noninvasive three-dimensional cardiac electrical imaging (3DCEI) vs. Intracavitary noncontact mapping (NCM) was evaluated on Localization error of the initiation site of activation. Noninvasive 3D cardiac electrical imaging accurately localized the initiation site of activation with an average error of 7.0+/-1.1 mm for RV and 6.6+/-1.9 mm for LV sites in a swine model.