Key result
Approximating the intrinsic ATP formula to calculate pulse numbers may prevent ATP-induced VT acceleration.
Why the study?
Does changing conventional ICD settings to intrinsic antitachycardia pacing-like settings help prevent VT acceleration in patients with cardiac failure?
Does changing conventional ICD settings to intrinsic antitachycardia pacing-like settings help prevent VT acceleration in patients with cardiac failure?
Answers are grounded in this paper and cite their sources.
Calculating the required number of ATP pulses using an iATP formula approximation may optimize conventional ICD programming and prevent VT acceleration.
Key Teaching Points•Using an approximation of the intrinsic antitachycardia pacing (iATP) formula, the number of antitachycardia pacing (ATP) pulses required to reach the ventricular tachycardia (VT) circuit (the number of pulses required for S1) can be calculated.•Determining the number of S1 pulses would help calculate whether the VT circuit can be reached with the set number of ATP pulses, even for patients using conventional implantable cardioverter-defibrillators and cases wherein iATP cannot be used.•Knowledge of the minimum number of ATP pulses necessary (up to the number of pulses in S1 + 2 pulses) may help prevent VT acceleration by ATP. •Using an approximation of the intrinsic antitachycardia pacing (iATP) formula, the number of antitachycardia pacing (ATP) pulses required to reach the ventricular tachycardia (VT) circuit (the number of pulses required for S1) can be calculated.•Determining the number of S1 pulses would help calculate whether the VT circuit can be reached with the set number of ATP pulses, even for patients using conventional implantable cardioverter-defibrillators and cases wherein iATP cannot be used.•Knowledge of the minimum number of ATP pulses necessary (up to the number of pulses in S1 + 2 pulses) may help prevent VT acceleration by ATP.
No takes yet. Share an insight, caveat, or question.
Tachibana et al. (2022) studied Ventricular tachycardia in patients with cardiac failure. Intrinsic antitachycardia pacing-like settings vs. Conventional implantable cardioverter-defibrillator settings was evaluated. Using an approximation of the intrinsic antitachycardia pacing formula to calculate the required number of ATP pulses may help prevent ventricular tachycardia acceleration by ATP.
Synapse has enriched 10 related papers. Consider them for comparative context: