Key result
Radial systolic BP was on average 5.5 mm Hg higher than brachial systolic BP, with only 43% of participants having radial systolic BP within ±5 mm Hg of brachial.
Why the study?
Differences between brachial and radial systolic BP have never been accurately quantified, despite radial intra-arterial BP being used as a reference standard for validating brachial cuff BP devices.
Does radial intra-arterial systolic blood pressure differ from brachial intra-arterial systolic blood pressure in patients undergoing coronary angiography?
Cross-Sectional (n=180)
Does radial intra-arterial systolic blood pressure differ from brachial intra-arterial systolic blood pressure in patients undergoing coronary angiography?
Radial systolic BP is not representative of brachial systolic BP, indicating that validation testing of BP devices utilizing intra-arterial BP as the reference standard should use measurements from the same site.
Radial-brachial systolic BP differences may require site-matched references for device validation; Level 4 data leaves open clinical translation.
Radial intra-arterial blood pressure (BP) is sometimes used as the reference standard for validation of brachial cuff BP devices. Moreover, there is an emerging wearables market seeking to measure BP at the wrist. However, radial systolic BP may differ when compared with brachial; yet some authors have labeled these differences as a fictional Popeye phenomenon. Indeed, differences between brachial and radial systolic BP have never been accurately quantified, and this was the aim of this study. Intra-arterial BP was measured consecutively at the brachial and radial artery in 180 participants undergoing coronary angiography (aged 61±10 years; 69% men). On average, radial systolic BP was 5.5 mm Hg higher than brachial systolic BP. Only 43% of participants had radial systolic BP within ±5 mm Hg of brachial. Additionally, 46%, 19%, and 13% of participants had radial systolic BP >5, between 5 and 10, and between 10 and 15 mm Hg higher than brachial respectively. A further 14% of participants had radial systolic BP >15 mm Hg higher than brachial, representing the so-called Popeye phenomenon. Finally, 11% of participants had radial systolic BP >5 mm Hg lower than brachial. In conclusion, radial systolic BP is not representative of brachial systolic BP, with most participants having a radial systolic BP >5 mm Hg higher than brachial and many with differences >15 mm Hg. Therefore, validation testing of BP devices utilizing intra-arterial BP as the reference standard should use intra-arterial BP measured at the same site as the brachial cuff or wearable device.
No takes yet. Share an insight, caveat, or question.
Armstrong et al. (2019) conducted a cross-sectional in Undergoing coronary angiography (n=180). Radial intra-arterial BP measurement vs. Brachial intra-arterial BP measurement was evaluated on Difference between brachial and radial systolic BP. Radial systolic BP was on average 5.5 mm Hg higher than brachial systolic BP, with only 43% of participants having radial systolic BP within ±5 mm Hg of brachial.