Key result
The Dimension Vista LOCI digoxin assay is virtually free from clinically significant interferences by spironolactone, potassium canrenoate, and canrenone, showing a maximum negative bias of 11.3%.
Why the study?
Does supplementation with spironolactone, potassium canrenoate, or canrenone interfere with the Dimension Vista LOCI digoxin assay in serum pools?
Does supplementation with spironolactone, potassium canrenoate, or canrenone interfere with the Dimension Vista LOCI digoxin assay in serum pools?
Answers are grounded in this paper and cite their sources.
The Dimension Vista LOCI digoxin assay is virtually free from clinically significant interferences by spironolactone, potassium canrenoate, and canrenone.
Spironolactone, a potassium-sparing diuretic metabolized to canrenone is often used with digoxin to treat various conditions including congestive heart failure. Potassium canrenoate is a similar drug, which is also metabolized to canrenone. Due to reported both positive and negative interference of spironolactone, potassium canrenoate, and their common metabolite canrenone with digoxin immunoassays, we investigated potential interference of these compounds with the new homogenous sequential chemiluminescent assay for digoxin based on the luminescent oxygen channeling technology (LOCI digoxin) for application on the Dimension and Vista platform. When aliquots of a drug-free serum pool were supplemented with various amounts of spironolactone, potassium canrenoate, or canrenone and apparent digoxin values were measured using Dimension Vista LOCI digoxin assay, we observed no detected value except when aliquots were supplemented with very high amounts of potassium canrenoate or canrenone. However, we observed that apparent digoxin concentrations were very low. When aliquots of a serum digoxin pool (prepared by pooling specimens from patients receiving digoxin), were further supplemented with various amounts of spironolactone, potassium canrenoate, or canrenone and serum digoxin concentrations were remeasured using the LOCIdigoxin assay, only statistically significant falsely lower digoxin values (negative interference) were observed in specimens containing very high amounts of canrenone or potassium canrenoate. However, such small bias may not have any clinical significance. We conclude that new Dimension Vista LOCI digoxin assay is virtually free from interferences of spironolactone, potassium canrenoate, and their common metabolite canrenone.
No takes yet. Share an insight, caveat, or question.
Dasgupta et al. (2012) studied Digoxin therapeutic drug monitoring. Spironolactone, potassium canrenoate, and canrenone vs. Unsupplemented digoxin serum pool was evaluated on Change in measured digoxin concentration with 2,000 ng/mL canrenone (-11.3% bias, p=<0.05). The Dimension Vista LOCI digoxin assay is virtually free from clinically significant interferences by spironolactone, potassium canrenoate, and canrenone, showing a maximum negative bias of 11.3%.