Key result
The ECLIA-Digoxin assay was virtually free from interference by spironolactone, potassium canrenoate, and canrenone when measuring digoxin concentrations in serum samples.
Why the study?
Does the ECLIA-Digoxin assay prevent interference from spironolactone and related compounds compared to FPIA?
Does the ECLIA-Digoxin assay prevent interference from spironolactone and related compounds compared to FPIA?
The ECLIA-Digoxin assay is virtually free from interference by spironolactone and its metabolites, improving the accuracy of digoxin monitoring in patients on these diuretics.
Answers are grounded in this paper and cite their sources.
May improve digoxin monitoring accuracy in spironolactone users; leaves open prospective validation versus other assays.
Spironolactone and potassium canrenoate (aldosterone antagonist diuretics) are sometimes used in conjunction with digoxin for patient management. Spironolactone, potassium canrenoate, and their common metabolite canrenone interfere with serum digoxin measurement using various immunoassays. Recently a new enzyme-linked chemiluminescent immunosorbent digoxin assay (ECLIA-Digoxin) became commercially available for application on the ADVIA IMS 800i modular system (Bayer HealthCare, Tarrytown, NY). We investigated the potential interference of spironolactone and related compounds in this assay by comparing the results with the fluorescence polarization immunoassay (FPIA), which is known to have significant cross-reactivity with these compounds as well as a turbidimetric assay for digoxin with no known cross-reactivity with spironolactone and related compounds. Aliquots of drug free serum were supplemented with therapeutic and above therapeutic concentrations of spironolactone, canrenone, and potassium canrenoate, and apparent digoxin concentrations were measured. No apparent digoxin concentration was observed using the ECLIA-Digoxin or turbidimetric assay. When serum pools prepared from patients receiving digoxin were further supplemented with these compounds, we observed no significant change in digoxin concentrations in the presence of these compounds with the ECLIA-Digoxin. We conclude that this assay is virtually free from interferences from spironolactone, potassium canrenoate and their common metabolite canrenone.
No takes yet. Share an insight, caveat, or question.
Dasgupta et al. (2006) studied Digoxin measurement interference. ECLIA-Digoxin assay vs. Fluorescence polarization immunoassay (FPIA) and turbidimetric assay was evaluated on Interference from spironolactone, potassium canrenoate, and canrenone. The ECLIA-Digoxin assay was virtually free from interference by spironolactone, potassium canrenoate, and canrenone when measuring digoxin concentrations in serum samples.