Highlights the importance of careful method selection in therapeutic drug monitoring to avoid inappropriate digoxin dosage adjustments due to assay interference.
Answers are grounded in this paper and cite their sources.
Potential assay interference warrants caution in digoxin TDM interpretation; case reports leave open the prevalence and mitigation of such discrepancies.
The authors present a case of an elderly female patient with heart failure and renal dysfunction treated with digoxin, where 2 commercial immunoassay methods (DRI, Microgenics, and DGNA, Dade Behring) showed a clinically very significant discrepancy on the same plasma sample, viz. 0.5 and 2.3 nmol/L, respectively. The sample was also referred to a third external laboratory that returned a result of 0.9 nmol/L using mFPIA (AxSYM, Abbott). Subsequent ultrafiltration (30,000 Dalton) on the sample essentially eliminated the difference, suggesting an interference from a large molecular weight compound(s), potentially the well-described digoxin-like immunoreactive substance(s) (DLIS). Although further study is required to verify that the DLIS implicated was indeed the interfering species, it does again highlight the importance of careful method selection in the clinical therapeutic drug monitoring laboratory to ensure that such well-established potential problems do not result in inappropriate dosage reduction with consequent lack of adequate drug exposure and serious clinical sequelae.