Key result
Diastolic CT-FFR outperforms systolic CT-FFR in detecting hemodynamically significant myocardial ischemia.
Why the study?
The diagnostic performance of machine learning-based CT-FFR between systolic and diastolic phases in identifying myocardial ischemia had not been compared against invasive FFR.
Does diastolic CT-FFR improve diagnostic performance for identifying myocardial ischemia compared to systolic CT-FFR in patients with coronary artery disease?
Observational (n=96)
Blinded
No
Does diastolic CT-FFR improve diagnostic performance for identifying myocardial ischemia compared to systolic CT-FFR in patients with coronary artery disease?
Machine learning-based diastolic CT-FFR provides superior diagnostic performance and narrower limits of agreement compared to systolic CT-FFR for identifying myocardial ischemia.
Answers are grounded in this paper and cite their sources.
Diastolic-phase CT-FFR may improve ischemia detection; leaves open whether phase selection changes management or outcomes.
CT-derived fractional flow reserve (CT-FFR) is a powerful tool for identifying hemodynamic ischemia. Coronary CT angiography (CCTA) images with the best quality in a cardiac cycle are conventionally reconstructed in clinical practice. To compare the diagnostic performance of machine learning (ML)-based CT-FFR between systolic and diastolic phases in identifying myocardial ischemia, using invasive fractional flow reserve (FFR) as a reference standard. From December 2020 to October 2021, consecutive coronary artery disease (CAD) patients who underwent coronary computed tomography angiography (CCTA) and invasive FFR were prospectively enrolled. Prospective electrocardiographic (ECG)-triggered scan from 30% to 80% of R-R interval was applied on image extraction of all the patients. CT-FFR was implemented in systolic and diastolic phases for each target vessel. Ninety-six patients (mean age 62 ± 8, 31.3% female) with 98 vessels were successfully simulated in both systolic and diastolic phases. The area under the curve (AUC) was significantly higher in diastolic CT-FFR (0.885 vs. 0.790, p < 0.001). The limits of agreement (LoA) range was narrower in diastolic CT-FFR (-0.134 to 0.148), compared with systolic CT-FFR (-0.234 to 0.273). Diastolic CT-FFR performed better than systolic CT-FFR in sensitivity (81.5% vs. 66.7%), specificity (83.1% vs. 71.8%), and accuracy (82.7% vs. 70.4%, all p < 0.05). The LoA range was narrower for diastolic CT-FFR across all stenosis degree groups, with diagnostic performance better in sensitivity (84.0% vs. 68.0%) and accuracy (81.4% vs. 58.1%, p < 0.001) among severe stenosis group. In ML-based CT-FFR methods, diastolic CT-FFR showed better diagnostic performance and narrower LoA than systolic CT-FFR. The study has been registered through the Ethics Committee of Zhongshan Hospital, Fudan University, with the clinical trial number B2020-088R. The registration date is May 15, 2020.
No takes yet. Share an insight, caveat, or question.
Guo et al. (2026) conducted an observational in Coronary artery disease (n=96). Diastolic CT-FFR vs. Systolic CT-FFR was evaluated on Area under the curve (AUC) for identifying hemodynamically significant ischemia (95% CI 0.816-0.955, p=<0.001). Diastolic CT-FFR improved the identification of hemodynamically significant myocardial ischemia compared to systolic CT-FFR, demonstrating a higher AUC.
Synapse has enriched 10 related papers. Consider them for comparative context: