Key result
MostCare-PRAM estimated Cardiac Index with a strong correlation (r = 0.93, p < 0.0001) and good agreement compared to the Fick method in pediatric patients undergoing cardiac surgery.
Why the study?
Does the MostCare-PRAM method accurately estimate cardiac index compared to the Fick method in pediatric patients undergoing cardiac surgery?
Observational (n=13)
No
Does the MostCare-PRAM method accurately estimate cardiac index compared to the Fick method in pediatric patients undergoing cardiac surgery?
Effect estimate: r = 0.93
Absolute Event Rate: 2.53% vs 2.51%
p-value: p=<0.0001
The MostCare-PRAM system provides a reliable, uncalibrated, minimally invasive estimation of cardiac index that shows good agreement with the reference Fick method in pediatric patients undergoing cardiac surgery.
May support MostCare-PRAM for CI monitoring in pediatric cardiac surgery; leaves open larger prospective validation.
Background: There is an increased interest in methods of objective cardiac output measurement in pediatric cardiac surgery. Several techniques are available, but have limitations, among the new technologies pressure recording analytical method with MostCare (MostCare-PRAM), a minimally invasive hemodynamic monitoring system, represents a novel arterial pulse contour method that does not require calibration. For this reason, we compared the MostCare-PRAM vs the Fick method for estimation of cardiac output. Methods: We studied prospectively 13 pediatric patients who underwent cardiac surgery and compared intraoperatively Cardiac Index (CI) measured with the MostCare-PRAM with the CI measured with the Fick method. We also measured Cardiac Cycle Efficiency (CCE) and maximal arterial pressure/time ratio (dp/dt max) and compared with Fick method. Results: The data showed good agreement between CI Fick and CI MostCare-PRAM (r = 0.93 and R 2 = 0.86; p < 0.0001) and also between CCE (r = 0.82 and R 2 = 0.67; p < 0.001) and dp/dt (r = 0.84; R 2 = 0.81; p < 0.001) with CI measured with Fick method. Conclusion: In pediatric patients submitted to cardiac surgery, the MostCare-PRAM seems to estimate CI with a good level of agreement with the Fick method measurements.
No takes yet. Share an insight, caveat, or question.
Pittarello et al. (2018) conducted an observational in Congenital heart disease (n=13). MostCare-PRAM vs. Fick method was evaluated on Cardiac Index (CI) correlation (r = 0.93, p=<0.0001). MostCare-PRAM estimated Cardiac Index with a strong correlation (r = 0.93, p < 0.0001) and good agreement compared to the Fick method in pediatric patients undergoing cardiac surgery.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: