Why the study?
Does the Pressure Recording Analytical Method (PRAM) accurately measure cardiac output compared to continuous thermodilution in patients with a pulsatile LVAD?
Does the Pressure Recording Analytical Method (PRAM) accurately measure cardiac output compared to continuous thermodilution in patients with a pulsatile LVAD?
PRAM provides accurate, uncalibrated continuous cardiac output monitoring from a peripheral artery in patients supported with a pulsatile LVAD.
Supports PRAM for continuous CO monitoring in pulsatile LVAD; hypothesis-generating, requires prospective validation before practice change.
We evaluated the accuracy of an uncalibrated pulse contour method called Pressure Recording Analytical Method (PRAM) compared with continuous thermodilution for cardiac output (CO) monitoring in patients implanted with a pulsatile left ventricular assist device (LVAD). Twelve adult patients implanted with the HeartMate I-XVE device were studied. CO was simultaneously evaluated by PRAM and by continuous thermodilution. Blood flow values displayed by the LVAD's console were also used for the comparison. Bland-Altman and linear regression analyses were applied. A total of 72 CO measurements (range 3.8-6.7 l/min) were obtained. Mean CO was 5.23±0.70 l/min for the 'hot' pulmonary thermodilution (ThD-CCO) method, 5.28±0.63 l/min for PRAM and 4.83±0.67 l/min for LVAD-CO. A high correlation (r=0.90), a good agreement (mean bias -0.04 l/min, precision ±0.38 l/min) and a low percentage of error (7.3%) were observed between PRAM-CO and ThD-CCO. A good correlation was found between LVAD-CO and either ThD-CCO (r=0.88) or PRAM-CO (r=0.86), but an overestimation of 10% was observed for both PRAM-CO (mean bias -0.44 l/min) and ThD-CCO (mean bias -0.40 l/min). Our results demonstrated good agreements between PRAM-CO, ThD-CCO and LVAD-CO. PRAM derives CO from a peripheral artery without calibration and may be a complementary tool in the hemodynamic assessment of patients supported with a VAD.
No takes yet. Share an insight, caveat, or question.
Scolletta et al. (2011) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: