Key result
MostCare-PRAM shows good agreement with the Fick method for estimating pediatric cardiac index.
Why the study?
Does MostCare-PRAM accurately estimate cardiac index compared to the Fick method in pediatric patients undergoing cardiac catheterization?
Observational (n=52)
No
Does MostCare-PRAM accurately estimate cardiac index compared to the Fick method in pediatric patients undergoing cardiac catheterization?
Effect estimate: r2 = 0.90
p-value: p=<0.001
The MostCare-PRAM method provides a reliable estimation of cardiac index in pediatric patients undergoing cardiac catheterization, showing good agreement with the reference Fick method.
PRAM may support continuous CI monitoring in stable pediatric cath patients; leaves open prospective validation in unstable or perioperative settings.
Summary Background Pressure recording analytical method ( PRAM ) is a novel, arterial pulse contour method for measuring cardiac output ( CO ). Validation studies of PRAM in children are few, and have shown contradictory results. The aim of the study was to compare the MostCare ® ‐ PRAM vs the Fick method of cardiac output estimation (reference method). Methods This is a single‐center, prospective observational study in 52 pediatric patients who underwent diagnostic right and left heart catheterization. Cardiac index ( CI ) measurements with the MostCare ® ‐ PRAM vs the Fick method were obtained under hemodynamically stable conditions. Results Forty CI measurements were performed. The data showed good agreement between CI F ick and CI PRAM : r 2 = 0.90 ( P < 0.001), mean bias −0.075, limits of agreement from −0.56 to 0.41. The percentage error was 17%. Comparable results were obtained for patients <20 kg ( r 2 = 0.87, P < 0.001), mean bias −0.135, limits of agreement from −0.35 to 0.62, percentage of error 17%. Conclusions In pediatric patients undergoing diagnostic right and left heart catheterization, the MostCare ® ‐ PRAM was shown to estimate CI with a good level of agreement with the Fick method measurements.
No takes yet. Share an insight, caveat, or question.
Alonso-Íñigo et al. (2016) conducted an observational in pediatric patients undergoing diagnostic right and left heart catheterization (n=52). MostCare-PRAM vs. Fick method was evaluated on Cardiac index (CI) measurement agreement (r2 = 0.90, p=<0.001). MostCare-PRAM showed good agreement with the Fick method for estimating cardiac index in pediatric patients undergoing cardiac catheterization (r2 = 0.90; P<0.001; mean bias -0.075).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: