Answers are grounded in this paper and cite their sources.
Demonstrates accurate determination of 99mTc-DTPA clearance in patients, indicating a potential routine method.
To derive a formula for determination of 99mTc-DTPA clearance (Cl) from the radioactivity in a single plasma sample, the relation Cl = ECV/t was used, where ECV is the extracellular volume and t is the mean transit time of the tracer in the organism. By studying 99mTc-DTPA time activity curves of 45 consecutive patients we found that ECV could be estimated from the body surface area, and that t could be calculated from the radioactivity in a single plasma sample. Cl calculated by the single sample method was almost identical to Cl calculated by a standard multiple sample method (r = 0.987). It is concluded that the single sample method is accurate and that it may prove useful as a routine method provided that the method is not used in patients with 99mTc-DTPA clearance less than 30 ml min-1, since this investigation includes only patients with Cl greater than 30 ml min-1; the method is not used for children, since the test material did not contain any children; the method is only used for plasma samples drawn at 180 less than or equal to t less than or equal to 300 min.
No takes yet. Share an insight, caveat, or question.
Christensen et al. (1986) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: