Key result
The Rossmax Mandaus II digital auscultatory device met AAMI criteria with a mean difference of -2.5+/-3.4 mmHg for SBP and -2.3+/-3.6 mmHg for DBP compared to a mercury sphygmomanometer.
Why the study?
Does the Rossmax Mandaus II provide accurate blood pressure measurements compared to a mercury sphygmomanometer in adults?
Observational (n=39)
Does the Rossmax Mandaus II provide accurate blood pressure measurements compared to a mercury sphygmomanometer in adults?
The Rossmax Mandaus II is the first auscultatory device with a numeric digital display to pass the ESH protocol, validating its accuracy for clinical use in adults.
Answers are grounded in this paper and cite their sources.
Supports device accuracy in adults; leaves open outcome trials before routine clinical adoption.
OBJECTIVE: To assess the accuracy of the Rossmax Mandaus II for clinical use, according to the protocol of the European Society of Hypertension (ESH). METHODS: Thirty-nine participants were recruited to fulfill the 33 participant requirement of the protocol. Nine sequential same-arm blood pressure readings were taken by trained observers alternating between simultaneous mercury sphygmomanometer readings and the device readings. The latter seven readings were analysed according to the protocol guidelines and the criteria of the Association for the Advancement of Medical Instrumentation. RESULTS: The device passed both phases of the ESH protocol. Out of 99 differences, 78 systolic (SBP) and 84 diastolic (DBP) differences were within 5 mmHg; all differences were within 10 and 15 mmHg for both SBP and DBP. In phase 2.2 for SBP and DBP, respectively, 28 of 33 and 31 of 33 participants had at least two out of their three differences within 5 mmHg. Only one participant was found to have all three differences greater than 5 mmHg for SBP and DBP, respectively. The device also fulfilled the criteria of the Association for the Advancement of Medical Instrumentation standard with a mean difference+/-SD of -2.5+/-3.4 mmHg (SBP) and -2.3+/-3.6 mmHg (DBP). CONCLUSION: The Rossmax Mandaus II is the first auscultatory device with a numeric digital display to pass the ESH protocol. It can be recommended for clinical use in an adult population.
No takes yet. Share an insight, caveat, or question.
Tasker et al. (2009) conducted an observational in Blood pressure measurement (n=39). Rossmax Mandaus II digital auscultatory device vs. Mercury sphygmomanometer was evaluated on Accuracy according to the European Society of Hypertension protocol and AAMI criteria (MD -2.5 mmHg (SBP) and -2.3 mmHg (DBP)). The Rossmax Mandaus II digital auscultatory device met AAMI criteria with a mean difference of -2.5+/-3.4 mmHg for SBP and -2.3+/-3.6 mmHg for DBP compared to a mercury sphygmomanometer.
Synapse has enriched 10 related papers. Consider them for comparative context: