August 24, 2006Blood Pressure Monitoring

Validation of three oscillometric blood pressure devices against auscultatory mercury sphygmomanometer in children

View Full Paper
Ask AI
Bookmark
Share

Key result

The Datascope Accutorr Plus passed the adapted IP-ESH validation for blood pressure measurement in children, whereas the Welch-Allyn Vital Sign Monitor and Dinamap Procare-120 failed.

Why the study?

Do the Welch-Allyn Vital Sign Monitor, Dinamap Procare-120, and Datascope Accutorr Plus accurately measure blood pressure compared to an auscultatory mercury sphygmomanometer in children aged 5-15 years?

Population

132 children aged 5-15 years old (67 boys, 65 girls)

Comparison

Blood pressure measurement using Welch-Allyn… vs Blood pressure measurement using auscultatory…

Design

Cross-sectional, Mercury sphygmomanometer read simultaneously by two…

Authors

SWSik‐Nin WongTuen Mun HospitalRSRita Yn Tz SungMichigan MedicineLLLettie Chuk‐Kwan LeungKwong Wah Hospital

Discussion

Loading...

Member takes

Implication

May warrant preference for Datascope Accutorr Plus in children; leaves open need for prospective validation trials.

Study Design

Type

Cross-Sectional (n=132)

Structured PICO

Do the Welch-Allyn Vital Sign Monitor, Dinamap Procare-120, and Datascope Accutorr Plus accurately measure blood pressure compared to an auscultatory mercury sphygmomanometer in children aged 5-15 years?

P
Population
132 children aged 5-15 years old (67 boys, 65 girls)
I
Intervention
Blood pressure measurement using Welch-Allyn Vital Sign Monitor, Dinamap Procare-120, or Datascope Accutorr Plus
C
Comparator
Blood pressure measurement using auscultatory mercury sphygmomanometer
O
Outcome
Validation according to the International Protocol of European Society of Hypertension adapted for children (mean difference and SD, percentage of readings within 5, 10, 15 mmHg)surrogate

The Datascope Accutorr Plus is a valid oscillometric blood pressure device for use in children aged 5-15 years, whereas the Welch-Allyn Vital Sign Monitor and Dinamap Procare-120 did not meet validation criteria.

Cite This Study

Wong et al. (2006) conducted a cross-sectional in Blood pressure measurement (n=132). Oscillometric BP devices (Welch-Allyn Vital Sign Monitor, Dinamap Procare-120, Datascope Accutorr Plus) vs. Auscultatory mercury sphygmomanometer was evaluated on Validation according to the adapted International Protocol of European Society of Hypertension. The Datascope Accutorr Plus passed the adapted IP-ESH validation for blood pressure measurement in children, whereas the Welch-Allyn Vital Sign Monitor and Dinamap Procare-120 failed.

synapsesocial.com/papers/6a159ca2cb0379474a826130https://doi.org/10.1097/01.mbp.0000209082.09623.b4

Topics

Hypertension management
View Full Paper
Ask AI
Bookmark
Share

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Evaluation of the Dinamap 8100 and Omron M1 blood pressure monitors for use in children2000 · 44 citations
  2. 2Differences in Blood Pressure Levels Obtained by Auscultatory and Oscillometric Methods1990 · 54 citations
  3. 3Validation of the Welch Allyn ‘Vital Signs’ blood pressure measurement device in pregnancy and pre‐eclampsia2003 · 37 citations
  4. 4Comparison of Auscultatory and Oscillometric Blood Pressures2001 · 216 citations