April 1, 2000Paediatric and Perinatal Epidemiology

Evaluation of the Dinamap 8100 and Omron M1 blood pressure monitors for use in children

View Full Paper
Ask AI
Bookmark
Share

Key result

The Dinamap 8100 overestimated systolic blood pressure by 11 mmHg (95% CI 9-12) and underestimated diastolic by 3 mmHg compared with a mercury sphygmomanometer in children.

Why the study?

Do the Dinamap 8100 and Omron M1 blood pressure monitors provide accurate measurements compared to a mercury sphygmomanometer in children?

Population

102 children recruited from infant and junior schools in Southampton

Comparison

Blood pressure measurement using Dinamap 8100 or… vs Blood pressure measurement using a mercury…

Design

Cross-sectional

Authors

MBMary BarkerUniversity of AgderASAlistair W. ShiellSouthampton General HospitalCLCatherine LawInstitute of Mental Health

Discussion

Loading...

Member takes

Implication

Limits clinical reliance on these monitors for pediatric BP; leaves open device validation needs in larger cohorts.

Study Design

Type

Cross-Sectional (n=102)

Multicenter

No

Structured PICO

Do the Dinamap 8100 and Omron M1 blood pressure monitors provide accurate measurements compared to a mercury sphygmomanometer in children?

P
Population
102 children recruited from infant and junior schools in Southampton underwent sequential blood pressure measurements to validate two automated monitors against a mercury sphygmomanometer.
E
Exposure
Blood pressure measurement using Dinamap 8100 (n=55) or Omron M1 (n=47)
C
Comparator
Blood pressure measurement using a mercury sphygmomanometer
O
Outcome
Accuracy of systolic and diastolic blood pressure readings compared to a mercury sphygmomanometersurrogate

Main Result

Mean Difference: 11 (95% CI 9–12)

Neither the Dinamap 8100 nor the Omron M1 blood pressure monitors meet British Hypertension Society accuracy criteria for clinical use in children.

Cite This Study

Barker et al. (2000) conducted a cross-sectional in Blood pressure measurement (n=102). Dinamap 8100 and Omron M1 blood pressure monitors vs. Mercury sphygmomanometer was evaluated on Difference in systolic blood pressure (Dinamap 8100 vs mercury sphygmomanometer) (MD 11 mmHg, 95% CI 9 to 12). The Dinamap 8100 overestimated systolic blood pressure by 11 mmHg (95% CI 9-12) and underestimated diastolic by 3 mmHg compared with a mercury sphygmomanometer in children.

synapsesocial.com/papers/6a5ac4909e3e30a63dac71cehttps://doi.org/10.1046/j.1365-3016.2000.00250.x

Topics

Hypertension management
View Full Paper
Ask AI
Bookmark
Share

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Inaccuracy of seven popular sphygmomanometers for home measurement of blood pressure1990 · 71 citations
  2. 2Blood pressure in first 10 years of life: the Brompton study.1992 · 107 citations
  3. 3Observer bias in blood pressure studies1988 · 145 citations
  4. 4Short report: An outline of the revised British Hypertension Society protocol for the evaluation of blood pressure measuring devices1993 · 522 citations
  5. 5The Dinamap 1846SX automated blood pressure recorder: comparison with the Hawksley random zero sphygmomanometer under field conditions.1992 · 69 citations