Key result
The A&D UA-778 blood pressure monitor showed a mean difference of 4.9 ± 6.3 mmHg for systolic and 2.4 ± 5.9 mmHg for diastolic pressure compared to a mercury sphygmomanometer, achieving a B/A grade.
Why the study?
Does the oscillometric blood pressure monitor UA-778 accurately measure blood pressure in children aged 4-15 years compared to a mercury sphygmomanometer?
Observational (n=85)
Does the oscillometric blood pressure monitor UA-778 accurately measure blood pressure in children aged 4-15 years compared to a mercury sphygmomanometer?
The UA-778 oscillometric blood pressure monitor is an accurate device for measuring blood pressure in children and adolescents.
Answers are grounded in this paper and cite their sources.
May support UA-778 use in children; leaves open broader device validation across populations.
OBJECTIVE: To validate the oscillometric blood pressure monitor UA-778 in children aged 4-15 years according to the British Hypertension Society protocol. METHOD: In 85 children (38 boys, 47 girls), nine sequential readings were taken simultaneously by two trained observers alternating between the mercury sphygmomanometer and the test device. The last seven readings were analyzed according to the British Hypertension Society protocol. The American Association for the Advancement of Medical Instrumentation criterion (mean difference +/- SD < 5 +/- 8 mmHg) was also applied. RESULTS: The difference between UA-778 and mercury sphygmomanometer readings was 4.9 +/- 6.3 mmHg for systolic pressure and 2.4 +/- 5.9 mmHg for diastolic pressure. The proportions of device-observer blood pressure differences within 5, 10, and 15 mmHg were 50, 83, and 95% for systolic blood pressure and 70, 90, and 96% for diastolic pressure (B/A grade according to the British Hypertension Society protocol). CONCLUSION: The oscillometric blood pressure monitor UA-778 seems to be an accurate device for blood pressure measurement in children and adolescents.
No takes yet. Share an insight, caveat, or question.
Нароган et al. (2009) reported an observational. A&D UA-778 blood pressure monitor vs. Mercury sphygmomanometer was evaluated on Difference between UA-778 and mercury sphygmomanometer readings (MD 4.9 +/- 6.3 mmHg (systolic) and 2.4 +/- 5.9 mmHg (diastolic)). The A&D UA-778 blood pressure monitor showed a mean difference of 4.9 ± 6.3 mmHg for systolic and 2.4 ± 5.9 mmHg for diastolic pressure compared to a mercury sphygmomanometer, achieving a B/A grade.
Synapse has enriched 10 related papers. Consider them for comparative context: