Key result
Tachycardia templates correctly discriminated 97% of ventricular depolarizations using correlation waveform analysis and 99% using area of difference during reinduction of monomorphic VT.
Why the study?
Can tachycardia templates constructed from previous episodes accurately recognize recurrent monomorphic ventricular tachycardia?
Observational (n=12)
Can tachycardia templates constructed from previous episodes accurately recognize recurrent monomorphic ventricular tachycardia?
Absolute Event Rate: 97% vs 99%
Tachycardia templates using correlation waveform analysis and area of difference can highly accurately discriminate recurrent monomorphic VT during acute electrophysiologic studies.
May support template-based recognition of recurrent monomorphic VT; hypothesis-generating from single case report.
The authors show that the ventricular electrograms of recurrent monomorphic ventricular tachycardia (VT) having an identical 12-lead surface configuration can be matched to a template constructed from a previous episode of the same VT. This could allow assessment of the number of times a device has delivered therapy for a known VT, in addition to the total number of therapies for cardiac rhythms that may have exceeded a rate criterion and/or morphology criterion. Templates constructed during monomorphic VT in 12 patients were examined by means of correlation waveform analysis (CWA) and area of difference (AD). With CWA the VT templates correctly discriminated 902/936 (97%) ventricular depolarizations, and AD correctly discriminated 932/936 (99%) depolarizations when the same VT was reinduced. The study used acute leads positioned during electrophysiologic study and has not been tested or validated on chronic leads. The monomorphic VT was reinduced during a single electrophysiology study. Whether changes in heart substrate or structure may cause changes in the intracardiac VT morphology over time has not been determined.< <ETX xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">></ETX>
No takes yet. Share an insight, caveat, or question.
Throne et al. (2003) conducted an observational in Monomorphic ventricular tachycardia (n=12). Correlation waveform analysis (CWA) vs. Area of difference (AD) was evaluated on Correct discrimination of ventricular depolarizations upon reinduction of the same VT. Tachycardia templates correctly discriminated 97% of ventricular depolarizations using correlation waveform analysis and 99% using area of difference during reinduction of monomorphic VT.