Clinicians should recognize that apparent AV interval extension in DDDR pacemakers with sensor-based algorithms during high atrial rates is a normal behavior, not a malfunction.
Answers are grounded in this paper and cite their sources.
Apparent AV interval extension may reflect normal DDDR sensor behavior, not malfunction; case report leaves open prevalence and management implications.
Rapid ventricular tracking response to supraventricular tachyarrhythmia is one major limitation to DDD pacing. In a DDDR pacemaker, sensor-based algorithms have been used to control these arrhythmias. These include the use of an interim rate limit (conditional ventricular tracking limit) or a separate maximum tracking and sensor rate limits (discrepant upper rate). These algorithms limit inappropriate ventricular pacing rate during tracking of pathological supraventricular tachyarrhythmia and atrial flutter by Wenckebach-like prolongation of the AV interval. We observed that this may cause an unexpected extension of the AV interval in patients with high atrial rate and intact AV nodal conduction. This was due to P wave rate above the conditional ventricular tracking limit or maximum tracking limit, but AV paced interval prolongation was avoided by the occurrence of intrinsic conduction, albeit at an AV interval longer than the programmed AV interval. This might appear as failure of ventricular pacing on the ECG. This phenomenon is a modified form of "upper rate" behavior occurring in the AV interval, and should be recognized as a normal behavior rather than pacemaker malfunction.