Key result
The Achieve Advance mapping catheter significantly increased the visualization rate of real-time pulmonary vein recordings compared to the previous Achieve catheter (73.5% vs 56.8%, p=0.0001) during cryoballoon ablation.
Why the study?
The new Achieve Advance mapping catheter includes a solid core designed to improve rotational response, but its rate of real-time pulmonary vein potential recording visualization compared with the previous catheter was unknown.
Does the new Achieve Advance mapping catheter improve the rate of visualisation of real-time recordings compared to the previous Achieve catheter during cryoballoon ablation of atrial fibrillation?
Population
200 consecutive patients undergoing CB ablation of AF
Comparison
Achieve Advance vs previous Achieve mapping catheter
Design
Comparative cohort study
Authors
Loading...
May facilitate real-time PV monitoring during cryoballoon ablation; leaves open impact on procedural success or safety.
Cohort (n=200)
Yes
Does the new Achieve Advance mapping catheter improve the rate of visualisation of real-time recordings compared to the previous Achieve catheter during cryoballoon ablation of atrial fibrillation?
Absolute Event Rate: 73.5% vs 56.8%
p-value: p=0.0001
The new Achieve Advance mapping catheter significantly improves the visualization of real-time pulmonary vein potentials during cryoballoon ablation compared to the previous generation.
Mugnai et al. (2019) conducted a cohort in Atrial fibrillation (n=200). Achieve Advance inner lumen mapping catheter vs. Previous Achieve mapping catheter was evaluated on Visualization of real-time recordings of pulmonary vein potentials (p=0.0001). The Achieve Advance mapping catheter significantly increased the visualization rate of real-time pulmonary vein recordings compared to the previous Achieve catheter (73.5% vs 56.8%, p=0.0001) during cryoballoon ablation.