Why the study?
Does continuous noninvasive tonometric blood pressure monitoring accurately reproduce intra-arterial blood pressure in normotensive and hypertensive subjects?
Does continuous noninvasive tonometric blood pressure monitoring accurately reproduce intra-arterial blood pressure in normotensive and hypertensive subjects?
The JENTOW continuous noninvasive tonometric blood pressure monitor is clinically acceptable and reliable for resting and slow-changing blood pressure measurements, though limited for rapid transient changes.
Answers are grounded in this paper and cite their sources.
Supports noninvasive tonometry for resting BP monitoring in research; leaves open validation during dynamic changes before clinical adoption.
A validation study of the continuous noninvasive tonometric blood pressure monitor called JENTOW was performed in 20 normotensive subjects and 10 hypertensive patients. Tonometric and intra-arterial blood pressures were simultaneously recorded at supine rest and during a Valsalva maneuver and tilting test. The results of the strict evaluation of the instrument's capacity for reproducing intra-arterial blood pressure were as follows: 1) The overall frequency response of the transcutaneous blood pressure-monitoring system based on arterial tonometry was flat, with negligible delay to intra-arterial blood pressure in the range of 0-5 Hz. 2) The largest discrepancy between intra-arterial and tonometric pressure waveforms was found at the early systolic phase; except for this phase, the tonometric waveform was almost equal to the intra-arterial waveform. 3) The beat-to-beat variability of tonometric pressure corresponded to that of intra-arterial pressure almost perfectly in the physiologically significant frequency range of 0-0.5 Hz. 4) During resting conditions, the averages of the systolic and diastolic values measured tonometrically corresponded well to those measured intra-arterially. 5) The changes in the between-method discrepancy of blood pressure values during the Valsalva maneuver were statistically significant but small (< 5 mm Hg). 6) No significant effect of postural tilting was found on the between-method discrepancy. We conclude that this method is clinically acceptable and reliable except for its limited capacity for recording the higher frequency intra-arterial waveform and for responding to the relatively rapid and large transient changes in blood pressure.
No takes yet. Share an insight, caveat, or question.
Sato et al. (1993) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: