Key result
Arterial tonometry provided accurate blood pressure monitoring during controlled hypotension, correlating well with intraarterial measurements.
Why the study?
Does arterial tonometry accurately measure blood pressure compared to intraarterial measurement in adult patients undergoing orthopedic surgery under controlled hypotension?
Observational (n=28)
Does arterial tonometry accurately measure blood pressure compared to intraarterial measurement in adult patients undergoing orthopedic surgery under controlled hypotension?
Arterial tonometry provides accurate and reliable real-time monitoring of blood pressure during controlled hypotension, comparable to invasive intraarterial measurement.
Answers are grounded in this paper and cite their sources.
Supports noninvasive BP monitoring in controlled hypotension; leaves open prospective validation before practice change.
A newly developed arterial tonometer enabled us to measure the blood pressure waveforms in addition to determining systolic and diastolic pressures noninvasively and continuously. Twenty-eight adult patients undergoing orthopedic surgery under controlled hypotension were studied. Systolic blood pressure was reduced to two-thirds of baseline values with an infusion of nitroglycerin during nitrous oxide/enflurane anesthesia. Intraarterial blood pressures were simultaneously measured in either the right or the left radial artery with a cannula and a Gould P23XL calibrated transducer; tonometric monitoring was performed on the contralateral radial artery using a Colin CBM-3000 instrument. The outputs of the two blood pressure measurement instruments were recorded for later data analysis. The shape of the tonometric pressure waveform was nearly identical to the waveform recorded intraarterially even during controlled hypotension. Regression analyses of 2039 paired tonometric and intraarterial blood pressure values during the hypotensive period showed good correlations (r = 0.78 for systolic, r = 0.81 for mean, and r = 0.70 for diastolic pressures). The accuracy of systolic, mean, and diastolic readings was from 4 to 7 mm Hg with negligible bias and did not differ significantly among six systolic, four mean, and four diastolic pressure groups. Our results indicate that arterial tonometry can provide accurate, reliable, and real-time monitoring of blood pressure even during controlled hypotension.
No takes yet. Share an insight, caveat, or question.
Kemmotsu et al. (1991) conducted an observational in orthopedic surgery under controlled hypotension (n=28). Arterial tonometry vs. Intraarterial blood pressure measurement was evaluated on Correlation of tonometric and intraarterial blood pressure values (r = 0.78 for systolic, r = 0.81 for mean, r = 0.70 for diastolic). Arterial tonometry provided accurate blood pressure monitoring during controlled hypotension, correlating well with intraarterial measurements.