Key result
A wearable multichannel-PPG framework incorporating acceleration signals and a heart rate tracking algorithm provided an average absolute error of 0.94 beats/m during high-intensity exercise.
Why the study?
Does a multichannel-PPG sensor framework using truncated SVD and acceleration signals improve the accuracy of heart rate estimation during intensive exercise compared to single-channel or other multichannel methods?
Does a multichannel-PPG sensor framework using truncated SVD and acceleration signals improve the accuracy of heart rate estimation during intensive exercise compared to single-channel or other multichannel methods?
A novel multichannel-PPG sensor framework using SVD and acceleration signals provides highly accurate heart rate estimation during intensive exercise.
Answers are grounded in this paper and cite their sources.
May enable accurate wearable HR monitoring during intense exercise; leaves open validation in larger clinical cohorts.
We propose a multichannel-PPG sensor that uses truncated singular value decomposition (SVD) for accurate estimation of heart rate (HR) during high-intensity exercise. We implemented a sensor to acquire nine channel signals simultaneously, and to enable real-time monitoring of PPG signals and HR. For denoising, we used SVD programmed in the microcontroller unit and showed that the pure pulse is concentrated mostly within the first largest singular value with the related two eigenvectors by extracting the significant feature components of the pulse signals. Our proposed sensor was tested in 32 subjects during exercise of various intensities: walking and medium- and high-intensity jogging. We compared our proposed method with the single-channel method, the multichannel weighting method, and the multichannel best signal selection. Furthermore, we incorporated acceleration signals and a HR tracking algorithm, into our proposed multichannel framework and evaluated its performance using the 12 data sets from the 2015 IEEE Signal Processing Cup Database. Our method showed the lowest absolute errors in HR estimation when compared to the single-channel, multichannel weighting, and multichannel best signal selection methods, irrespective of exercise intensity. We also found that our proposed multichannel framework incorporating acceleration signals and a HR tracking algorithm provided the lowest average absolute error of 0.94 beats/m. The results show that our proposed multichannel framework can be used for accurate HR estimation during high-intensity exercise.
No takes yet. Share an insight, caveat, or question.
Lee et al. (2018) studied Heart rate monitoring during intensive exercise (n=32). Multichannel-PPG framework incorporating acceleration signals and a HR tracking algorithm vs. Single-channel method, multichannel weighting method, and multichannel best signal selection was evaluated on Average absolute error in heart rate estimation. A wearable multichannel-PPG framework incorporating acceleration signals and a heart rate tracking algorithm provided an average absolute error of 0.94 beats/m during high-intensity exercise.