Key result
Repeated cardiopulmonary exercise testing in healthy biologic controls over 1 year demonstrated a mean coefficient of variation of 6.59% for VO2, 6.41% for VCO2, and 6.32% for VE.
Why the study?
Although guidelines recommend biologic control testing to assess cardiopulmonary exercise testing instruments for quality control, minimal data exist on how to perform it.
Observational (n=4)
No
Long-term variability in steady-state CPET measurements demonstrates a mean coefficient of variation of ~6.3-6.6%, providing a practical ± 1.96 z-score threshold for biologic control quality assessment.
Answers are grounded in this paper and cite their sources.
May inform CPET quality thresholds in controls; leaves open applicability to patients and clinical use.
BACKGROUND: Cardiopulmonary exercise testing is an increasingly common test and is considered the accepted standard for assessing exercise capacity. Quantifying variability is important to assess the instrument for quality control purposes. Though guidelines recommend biologic control testing, there are minimal data on how to do it. We sought to describe variability for oxygen consumption (V̇ O 2 ), carbon dioxide production (V̇ CO 2 ), and minute ventilation (V̇ E ) at various work rates under steady-state conditions in multiple subjects over a 1-y period to provide a practical approach to assess and perform biologic control testing. METHODS: We performed a single-center, prospective study with 4 healthy subjects, 2 men and 2 women. Subjects performed constant work rate exercise tests for 6 min each at 25–100 W intervals on a computer-controlled cycle ergometer. Data were averaged over the last 120 s at each work rate to reflect stepwise steady-state conditions. Descriptive statistics, including the mean, median, range, SD, and coefficient of variation (CoV) are reported for each individual across the 4 work rates and all repetitions. As these data were normative, z-scores were utilized, and a value greater than ± 1.96 z-scores was used to define significant test variability. RESULTS: Subjects performed 16–39 biocontrol studies over 1-y. The mean CoV for all subjects in V̇ O 2 was 6.59%, V̇ CO 2 was 6.41%, and V̇ E was 6.32%. The ± 1.96 z-scores corresponded to a 9.4–18.1% change in V̇ O 2 , a 9.6–18.1% change in V̇ CO 2 , and a 9–21.5% change in V̇ E across the 4 workloads. CONCLUSIONS: We report long-term variability for steady-state measurement of V̇ O 2 , V̇ CO 2 , and V̇ E obtained during biocontrol testing. Utilizing ± 1.96 z-scores allows one to determine if a result exceeds expected variability, which may warrant investigation of the instrument.
No takes yet. Share an insight, caveat, or question.
DeCato et al. (2022) conducted an observational in Healthy (Biologic Controls) (n=4). Repeated cardiopulmonary exercise testing (biocontrol testing) was evaluated on Mean coefficient of variation (CoV) for oxygen consumption (VO2), carbon dioxide production (VCO2), and minute ventilation (VE). Repeated cardiopulmonary exercise testing in healthy biologic controls over 1 year demonstrated a mean coefficient of variation of 6.59% for VO2, 6.41% for VCO2, and 6.32% for VE.