Key result
Disabling pacemaker sleep mode suppresses nocturnal bradycardia and PMVT after AV node ablation.
Case Report (n=1)
In patients with AVN ablation, pacemaker 'sleep mode' can cause relative bradycardia and promote 'R on T' mediated polymorphic ventricular tachycardia, highlighting the need for careful programming.
Answers are grounded in this paper and cite their sources.
May warrant caution with sleep mode programming after AVN ablation; case report leaves open need for systematic study.
BACKGROUND: Atrioventricular node (AVN) ablation preceded by pacemaker insertion is a safe procedure used for both rate and symptom control in symptomatic atrial fibrillation. CASE SUMMARY: This is a case of bradycardia-induced PMVT which occurred after AVN ablation, despite the presence of a VVI-R pacemaker inserted a few months prior. Lower pacing rates were allowed during the night due to 'sleep mode' being enabled. The nocturnal bradycardia resulted in ventricular escape rhythm with frequent ectopy, leading to 'R on T' mediated PMVT. Turning the sleep function off suppressed the ectopic activity and prevented further PMVT. DISCUSSION: PMVT following AVN ablation has been well described in literature. Be it after AVN ablation, or when changing pacemaker programming, the common denominator is a relative bradycardic state which promotes ventricular ectopy and "R on T".
No takes yet. Share an insight, caveat, or question.
Firescu et al. (2025) conducted a case report in Symptomatic atrial fibrillation (n=1). Pacemaker sleep mode vs. Disabling sleep mode was evaluated on Polymorphic ventricular tachycardia (PMVT). Pacemaker sleep mode after atrioventricular node ablation induced nocturnal bradycardia and polymorphic ventricular tachycardia, which was suppressed by disabling the sleep function.
Synapse has enriched 10 related papers. Consider them for comparative context: