Key result
Stent implantation induced coronary artery spasm in 54% of swine, with higher rates associated with oversized stents and smaller arteries.
Why the study?
Does coronary artery stent implantation induce spasm, and what factors influence its development in a swine model?
Does coronary artery stent implantation induce spasm, and what factors influence its development in a swine model?
Coronary artery stent implantation frequently induces spasm in a swine model, highlighting the importance of judicious stent sizing to avoid overdistention, especially in smaller arteries.
Answers are grounded in this paper and cite their sources.
Stent oversizing may increase spasm risk in small vessels; leaves open whether findings translate to human PCI.
Mechanical stimulation is a known cause of arterial spasm. Since stent implantation in the coronary arteries imparts mechanical trauma, we investigated, in an animal model that simulates human coronary anatomy and physiology, the frequency of occurrence and factors that influence the development of spasm. In 28 Hanford miniature swine, 9 of which were atherosclerotic and 19 of which were nonatherosclerotic, spasm occurred in 15 (54%). Stent‐induced spasm directly caused the death of one animal and was implicated in the death of two others. Factors that significantly contributed to coronary spasm were the stent: artery ratio (more spasm with oversized stents) and the size of the artery (more spasm with smaller arteries). There was a trend toward more spasm in atherosclerotic arteries. Spasm occurred despite the use of antispasmodic agents before, during, and after the scenting procedure. Our data suggest that coronary artery spasm may be an important clinical accompaniment to stent implantation in man and may be minimized by the judicious selection of stent size to avoid overdistention of the artery and the avoidance of stent placement in smaller arteries . (J Interven Cardiol 1993; 6:149–155)
No takes yet. Share an insight, caveat, or question.
Rodgers et al. (1993) studied Coronary artery spasm (n=28). Stent implantation was evaluated on Occurrence of coronary artery spasm. Stent implantation induced coronary artery spasm in 54% of swine, with higher rates associated with oversized stents and smaller arteries.