Key result
The Tensioday ambulatory blood pressure monitor achieved an A grade and satisfied AAMI criteria with a mean difference of 1.4 +/- 5.3 mmHg for systolic and 1.0 +/- 4.7 mmHg for diastolic pressure.
Why the study?
Does the Tensioday ambulatory blood pressure monitor accurately measure blood pressure compared to a standard mercury sphygmomanometer?
Observational (n=85)
Does the Tensioday ambulatory blood pressure monitor accurately measure blood pressure compared to a standard mercury sphygmomanometer?
Mean Difference: 1.4
The Tensioday ambulatory blood pressure monitor meets BHS and AAMI accuracy criteria and can be recommended for clinical use.
No takes yet. Share an insight, caveat, or question.
Supports Tensioday accuracy in controlled validation; leaves open generalizability to diverse real-world ambulatory use.
Németh et al. (2002) conducted an observational in Blood pressure measurement (n=85). Tensioday oscillometric ambulatory blood pressure monitor vs. Standard mercury sphygmomanometer was evaluated on Accuracy of blood pressure measurement (mean difference between device and observers) (MD 1.4 +/- 5.3 mmHg (systolic) and 1.0 +/- 4.7 mmHg (diastolic)). The Tensioday ambulatory blood pressure monitor achieved an A grade and satisfied AAMI criteria with a mean difference of 1.4 +/- 5.3 mmHg for systolic and 1.0 +/- 4.7 mmHg for diastolic pressure.