Key result
PulseSelect matches FARAPULSE for acute PVI but is linked to higher cardiac biomarker release.
Why the study?
Randomized trials have demonstrated safety and efficacy of PFA for AF using the FARAPULSE system, but implementation of the PulseSelect system in an experienced center required evaluation.
Does pulsed field ablation with PulseSelect compared to FARAPULSE alter procedural outcomes and biomarker release in patients undergoing first pulmonary vein isolation for atrial fibrillation?
Observational (n=150)
No
Does pulsed field ablation with PulseSelect compared to FARAPULSE alter procedural outcomes and biomarker release in patients undergoing first pulmonary vein isolation for atrial fibrillation?
Answers are grounded in this paper and cite their sources.
PulseSelect PFA is feasible but associated with higher myocardial injury biomarkers and longer procedural times compared to FARAPULSE, though FARAPULSE causes more hemolysis.
Background Randomized trials have demonstrated safety and efficacy of pulsed field ablation (PFA) for atrial fibrillation (AF) using the FARAPULSE™ system (PFA F). We evaluated implementation of the PulseSelect™ system (PFA P) in a PFA-F experienced center. Methods This single-center study included 150 consecutive patients undergoing first pulmonary vein isolation (PVI) with PFA-P ( n = 75) or PFA-F (n = 75). Procedural characteristics, acute efficacy, safety endpoints, AF recurrence rate, myocardial, neural and hemolysis biomarkers (high-sensitive troponin T, creatine kinase (CK), creatine kinase MB, bilirubin, haptoglobin, lactate dehydrogenase and S100) post ablation were analyzed and compared. Results Baseline characteristics were comparable, except for a higher proportion of males in the PFA-F group. Acute PVI was comparable between groups (98% vs. 100%). Skin-to-skin time and radiation dose were similar, while fluoroscopy (14.3 vs. 11.9 min, p = 0.009) and LA-dwell time (18 vs. 14 min, p < 0.001) were higher in PFA P. No major complication occurred, while minor complication rates were comparable. Cardiac biomarker release was significantly higher in PFA-P (ΔCK: 203 vs. 127 U/l , p = 0.005; Δtroponin: 1212 vs. 1014 pg/ml, p = 0.035 (day 1)), whereas neural injury markers were similar. Hemolysis markers indicated greater changes with PFA-F (Δbilirubin: 0.3 vs. 0.2 mg/dl, p = 0.004; Δhaptoglobin: −46.5 vs. −27.5 mg/dl, p = 0.001 (day1)). Procedural parameters improved with increasing PFA-P experience. Conclusion Implementation of PFA-P is feasible with an evident learning curve in a PFA-F experienced center. Higher post-procedural cardiac biomarker levels may suggest a greater extent of local myocardial tissue damage after PFA P. The impact of biomarkers on long-term outcome requires further studies.
No takes yet. Share an insight, caveat, or question.
Gold et al. (2026) conducted an observational in Atrial fibrillation (n=150). PulseSelect system (PFA-P) vs. FARAPULSE system (PFA-F) was evaluated on Acute pulmonary vein isolation (PVI). PulseSelect pulsed field ablation achieved similar acute pulmonary vein isolation to FARAPULSE but resulted in higher cardiac biomarker release.