Key result
Valve-in-MHV increases mean pressure gradient by ~4 mmHg compared to standalone mechanical valves.
Why the study?
To evaluate the haemodynamic performance of a novel hybrid Valve-in-MHV configuration with a CoreValve transcatheter heart valve implanted within a composite valved graft annulus after leaflet removal.
Does a Valve-in-MHV configuration provide comparable haemodynamic performance to standalone THV or MHV in an in vitro model?
Does a Valve-in-MHV configuration provide comparable haemodynamic performance to standalone THV or MHV in an in vitro model?
Answers are grounded in this paper and cite their sources.
An in vitro model of Valve-in-MHV demonstrated elevated transvalvular pressure gradients and increased turbulence compared to standalone valves, suggesting a potential for higher thrombogenicity that necessitates careful patient selection and device sizing.
OBJECTIVES: This study evaluates the haemodynamic performance of a novel hybrid configuration-Valve-in-MHV-where a CoreValve transcatheter heart valve is implanted within the annulus of a composite valved graft after leaflet removal. METHODS: In vitro testing was performed using left heart mock loop combined with backlight particle image velocimetry. Three configurations were assessed: (i) mechanical heart valve (MHV), (ii) CoreValve THV, and (iii) Valve-in-MHV. Flow parameters were measured at cardiac outputs of 3 and 5 L/min. RESULTS: At 5 L/min, the Valve-in-MHV showed the highest PG (15.5 mmHg) and TKE (0.53 m2/s2), compared to the THV (10.9 mmHg, 0.31 m2/s2) and MHV (11.5 mmHg, 0.26 m2/s2). Effective orifice area was smallest for the Valve-in-MHV (1.44 cm2). The Valve-in-MHV generated a more physiological central jet than the MHV, but with increased turbulence and higher peak velocities (up to 2.66 m/s). CONCLUSIONS: Examining the mechanistic implications of Valve-in-MHV may offer valuable insights into the likelihood of adverse effects such as leaflet thrombosis and the development of pronounced pressure gradients in patients who are candidates for Valve-in-MHV.
No takes yet. Share an insight, caveat, or question.
Catalano et al. (2025) studied Mechanical heart valve failure. Valve-in-MHV configuration (CoreValve THV within SJM mechanical heart valve annulus) vs. Standalone mechanical heart valve (MHV) and CoreValve THV was evaluated on Mean transvalvular pressure gradient (PG) at cardiac output of 5 L/min (p=<0.001).
Synapse has enriched 10 related papers. Consider them for comparative context: