Key result
The subpectoral approach for cardiovascular implantable electronic device implantation resulted in a complication rate of 6.3% compared to 2.6% with the conventional prepectoral approach (p=0.25).
Why the study?
The standard prepectoral CIED implantation approach cannot be performed in patients with thin skin or cosmetic concerns, prompting evaluation of the feasibility and safety of the subpectoral approach.
Does the subpectoral approach for CIED implantation maintain safety and lead performance while improving cosmetic outcomes compared to the conventional prepectoral approach?
Cohort (n=266)
No
Does the subpectoral approach for CIED implantation maintain safety and lead performance while improving cosmetic outcomes compared to the conventional prepectoral approach?
Absolute Event Rate: 6.3% vs 2.6%
p-value: p=0.25
The subpectoral approach for CIED implantation is a feasible, safe, and cosmetically superior alternative to the conventional prepectoral approach, particularly for younger patients or those with cosmetic concerns.
May offer alternative for thin-skinned patients; leaves open need for randomized confirmation of safety.
BACKGROUND: The prepectoral implantation technique has been the standard procedure for cardiovascular implantable electronic device (CIED). However, it cannot be performed in such patients with thin skin or patients with cosmetic concerns. This study was designed to demonstrate the feasibility and safety of the subpectoral compared to the prepectoral approach. METHODS: We conducted a retrospective, nonrandomized comparison of the prepectoral (234 cases) and subpectoral approach (32 cases) in patients who received CIED implantation at a tertiary center between July 2012 and May 2015. We compared lead characteristics, procedure time and complications between the subpectoral and prepectoral approach. RESULTS: =0.06). In lead characteristics, there were no significant differences between the two groups. CONCLUSION: The subpectoral approach is technically feasible and non-inferior to the prepectoral approach, in the aspect of complication and lead characteristics, but seemed to take more procedure time. The subpectoral approach is a more reasonable choice for selected patients in whom the prepectoral approach is not feasible or in individuals who have cosmetic concerns.
No takes yet. Share an insight, caveat, or question.
Kim et al. (2019) conducted a cohort in Cardiovascular implantable electronic device (CIED) implantation (n=266). Subpectoral implantation approach vs. Prepectoral implantation approach was evaluated on Complications requiring surgical or medical intervention (p=0.25). The subpectoral approach for cardiovascular implantable electronic device implantation resulted in a complication rate of 6.3% compared to 2.6% with the conventional prepectoral approach (p=0.25).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: