Key result
Low baseline intrathoracic impedance linked to ~38% higher all-cause mortality.
Why the study?
Does directly measured baseline intrathoracic impedance predict all-cause mortality in patients with implanted ICD or CRT-D devices?
Cohort (n=146,238)
Yes
Does directly measured baseline intrathoracic impedance predict all-cause mortality in patients with implanted ICD or CRT-D devices?
Directly measured intrathoracic impedance independently predicts all-cause mortality in device patients and adds prognostic value to established fluid index thresholds.
May refine mortality risk assessment in ICD/CRT-D patients; leaves open whether impedance-guided management improves outcomes.
BACKGROUND: Intrathoracic impedance-derived OptiVol fluid index calculated using implanted devices has been shown to predict mortality; direct measurements of impedance have not been examined. We hypothesized that baseline measured impedance predicts all-cause mortality; changes in measured impedance result in a change in the predicted mortality; and the prognostic value of measured impedance is additive to the calculated OptiVol fluid index. METHODS AND RESULTS: A retrospective analysis of 146,238 patients within the Medtronic CareLink database with implanted devices was performed. Baseline measured impedance was determined using daily values averaged from month 6 to 9 after implant and were used to divide patients into tertiles: group L = low impedance, ≤ 65 ohms; group M = medium impedance, 66 to 72 ohms; group H = high impedance, ≥ 73 ohms. Change in measured impedance was determined from values averaged from month 9 to 12 post implant compared with the 6- to 9-month values. OptiVol fluid index was calculated using published methods. All-cause mortality was assessed beginning 9 months post implant; changes in mortality was assessed beginning 12 months post implant. Baseline measured impedance predicted all-cause mortality; 5-year mortality for group L was 41%, M was 29%, and H was 25%, P < 0.001 among all groups. Changes in measured impedance resulted in a change in the predicted mortality; the prognostic value of measured impedance was additive to the OptiVol fluid index. CONCLUSIONS: Direct measurements of intrathoracic impedance using an implanted device can be used to stratify patients at varying mortality risk.
No takes yet. Share an insight, caveat, or question.
Zile et al. (2015) conducted a cohort in Heart failure with implanted ICD or CRT-D devices (n=146,238). Low baseline intrathoracic impedance (≤ 65 ohms) vs. Medium intrathoracic impedance (66-72 ohms) was evaluated on All-cause mortality (HR 1.38, 95% CI 1.33-1.42, p=<0.001). Low baseline intrathoracic impedance (≤ 65 ohms) was associated with a higher risk of all-cause mortality compared to medium impedance (66-72 ohms) with an adjusted HR of 1.38.