Key result
Applying PEEP during EV-ICD introducer removal maintains normal shock impedance by preventing mediastinal air entrapment.
Why the study?
Elevated shock impedance during EV-ICD implantation may preclude DFT testing despite appropriate lead position, prompting investigation into reversible mechanisms and preventive strategies.
Does applying positive end-expiratory pressure during introducer removal prevent elevated shock impedance and facilitate immediate DFT testing in patients undergoing EV-ICD implantation?
Case Report (n=7)
Does applying positive end-expiratory pressure during introducer removal prevent elevated shock impedance and facilitate immediate DFT testing in patients undergoing EV-ICD implantation?
A simple procedural modification using positive end-expiratory pressure during EV-ICD implantation may prevent mediastinal air entrapment and subsequent elevated shock impedance, facilitating reliable intraoperative DFT testing.
Answers are grounded in this paper and cite their sources.
May facilitate immediate DFT testing; hypothesis-generating for mediastinal air management in EV-ICD.
Introduction Elevated shock impedance during extravascular implantable cardioverter‐defibrillator (EV‐ICD) implantation may preclude defibrillation threshold (DFT) testing despite appropriate retrosternal lead position. We describe suspected mediastinal air entrapment as a reversible mechanism and a simple preventive strategy. Methods and Results During early EV‐ICD implantation experience, two patients developed shock impedance > 200 Ω immediately after retrosternal lead placement, with associated sensing abnormalities. Lead repositioning did not normalize impedance, no pneumothorax was identified, and DFT testing was deferred. Shock impedance normalized spontaneously within 1–2 days, permitting successful DFT testing without lead revision. Transient mediastinal air entrapment along the defibrillation vector was suspected. A procedural modification was adopted in which 5–10 cm H 2 O positive end‐expiratory pressure was applied immediately before retrosternal introducer removal, followed by prompt EV‐ICD lead advancement. Among five consecutive subsequent implantations using this approach, initial shock impedance remained within the expected range, permitting immediate DFT testing without lead repositioning or delay. Conclusion Transient mediastinal air entrapment may cause elevated shock impedance during EV‐ICD implantation. Positive end‐expiratory pressure during introducer removal may mitigate this complication and facilitate reliable intraoperative DFT testing.
No takes yet. Share an insight, caveat, or question.
Kerley et al. (2026) conducted a case report in Extravascular implantable cardioverter-defibrillator (EV-ICD) implantation (n=7). Positive end-expiratory pressure during introducer removal vs. Standard introducer removal was evaluated on Initial shock impedance within expected range permitting immediate DFT testing. Applying 5-10 cm H2O positive end-expiratory pressure during introducer removal in 5 EV-ICD implantations maintained normal shock impedance, preventing suspected mediastinal air entrapment.