Key result
Five cases of early mitral bileaflet prosthesis dysfunction due to a stuck leaflet in a clean prosthesis were identified, highlighting causes other than thrombosis and the importance of intraoperative TEE.
Case Report (n=5)
A stuck leaflet in a clean mitral bileaflet prosthesis can occur early without thrombosis and may evolve without major symptoms, highlighting the importance of intraoperative TEE.
Answers are grounded in this paper and cite their sources.
Alerts clinicians to non-thrombotic causes of early mitral prosthesis dysfunction; leaves open validation of diagnostic strategies in larger cohorts.
OBJECTIVES: We present five cases of mitral bileaflet prosthesis dysfunction as a result of a stuck leaflet in closed position, diagnosed at surgery or during the first postoperative year. METHODS AND RESULTS: The diagnosis was made by echocardiography and could be confirmed by fluoroscopy in three patients. All cases had in common an early occurrence and the presence of a clean prosthesis by transesophageal echocardiography (TEE). In one patient the diagnosis was made by intraoperative TEE, and the problem could be solved by rotation of the occluding device. In the other four patients the diagnosis was made postoperatively and different strategies of treatment were taken. All but one case were reoperations, one had a dysfunctional biological prosthesis and three had been submitted to mitral valvoplasty (two rheumatic and one degenerative valves). According to our policy, patients started sodium warfarin therapy 24 h after surgery in order to keep International normalized ratio (INR) values between 2.5-3.5. Only in one case the presence of thrombus/pannus could be confirmed. CONCLUSIONS: We call the attention to causes other than thrombosis that can provoke leaflet block and to the importance of performing intraoperative TEE in patients submitted to mitral valve replacement. We also emphasize the fact that a stuck leaflet in a clean prosthesis can evolve without major symptoms or cardiac events.
No takes yet. Share an insight, caveat, or question.
Almeida et al. (2007) conducted a case report in Mitral bileaflet prosthesis dysfunction (stuck leaflet) (n=5). Mitral valve replacement with bileaflet prosthesis was evaluated on Diagnosis and management of stuck leaflet. Five cases of early mitral bileaflet prosthesis dysfunction due to a stuck leaflet in a clean prosthesis were identified, highlighting causes other than thrombosis and the importance of intraoperative TEE.
Synapse has enriched 10 related papers. Consider them for comparative context: