Key result
Lattice-tip catheter RF ablation near coronary arteries triggers transient chest pain and ST depression from spasm.
Why the study?
Achieving durable transmural lesions in ventricular arrhythmia ablation is challenging near coronary arteries, where pulsed field ablation is often linked to coronary artery spasm and thermal ablation rarely causes injury.
Case Report (n=1)
Ablation of RVOT arrhythmias using large-footprint lattice-tip catheters carries a risk of coronary artery spasm, but using gradual radiofrequency delivery allows for early detection and prompt cessation, preventing permanent coronary injury.
Answers are grounded in this paper and cite their sources.
Transient spasm risk with lattice-tip RF near coronaries warrants vigilance; case report leaves open optimal energy titration.
In ventricular arrhythmias (VAs) ablation, achievement of durable transmural lesions is critical in ensuring procedural success. However, it remains challenging when substrate is localized in certain regions (i.e. septum, epicardium)1 or near coronary arteries (CAs), for instance in the left ventricular (LV) summit2. New techniques, such as pulsed field ablation (PFA), have been proposed as an alternative to conventional radiofrequency ablation (RFA) for such cases. However, PFA application in proximity to CAs is frequently associated with severe CA spasm (CAS)3, whereas CAS or injury (CAI) is rarely described as a complication of thermal ablation procedures4,5.
No takes yet. Share an insight, caveat, or question.
Dell’Aquila et al. (2025) conducted a case report in Premature ventricular complexes (n=1). Radiofrequency ablation using a large-footprint dual-energy lattice-tip catheter (Sphere-9) was evaluated. Radiofrequency ablation of premature ventricular complexes using a large-footprint lattice-tip catheter near the coronary arteries caused transient chest pain and ST-segment depression likely due to coronary artery spasm.